Mechanical System

Problem 

 

 

        

Possible Cause   

 

   Solution

Having trouble focusing when 

using high magnification objective.

Objective touches cover glass while 

turning the nosepiece

Coarse Focusing Knob is too tight

Stage drifts and will not stay in the 

focal plane

Slide does not move smoothly

The image moves considerably 

when touching the stage

Cover glass is facing down.

Cover glass is too thick.

Cover glass faces down

Cover glass is too thick

Tension Adjustment Knob is too tight

Tension adjustment knob is too loose

The slide is not positioned correctly

The moveable specimen holder is not 

fixed properly

The stage is fashioned incorrectly

Make sure cover glass is facing up.

Use a standard cover glass with a thickness 

of 0.13-0.17mm

Turn the nosepiece into the right position until 

you hear it click into place

Clean the appropriate lenses

Loosen it to an appropriate position

Tighten it to an appropriate position

Adjust it correctly

Adjust it correctly

Fasten the stage correctly

 

Electrical System

Problem 

 

 

        

Possible Cause   

 

   Solution

Bulb does not work

Bulb burns out quickly

The field of view is not bright 

enough

No power supply

Bulb is not installed correctly

Bulb is burnt out

Wrong bulb is being used

Wrong bulb being used

Light adjustment knob is not being 

used correctly

Check connection of power cable

Install bulb correctly

Replace bulb

Replace with correct bulb

Replace with correct bulb

Adjust knob correctly

Summary of Contents for UX1B

Page 1: ...Richter Optica info richter optica com Instructions for Models UX1B Binocular and UX1T Trinocular Laboratory and University Biological Microscope...

Page 2: ...cular Eyepiece Tube Stage Specimen Holder Objectives Condenser Focusing Knob Coarse Focus Knob Fine Focus Knob Illumination on off switch Light Intensity Control Stage Y Axis Travel Knob Stage X Axis...

Page 3: ...fingerprints or oil Damage of lens surfaces occur when they come in contact with such contaminants and image quality is reduced BEFORE USE When moving the microscope use both hands and only lay on fl...

Page 4: ...he silver plate on the base of the microscope To access replace the batteries remove all screws in the corners of the silver metal plate Once screws are removed lift off metal plate and access battery...

Page 5: ...objective While looking through the eyepieces move the eyetubes either further apart or closer together until the right and left field of view become one Line up the dots 1 on either side of the basep...

Page 6: ...o the left closed For the objectives in between adjust the lever betwen these settings 8 Using the oil immersion objective a First use the 4x objective to focus the specimen b Place a drop of oil on t...

Page 7: ...agm closed too far No cover glass Too thick or thin cover glass Immersion oil not used on immersion procedure Air bubbles in immersion oil Immersion oil used on a dry objective Greasy residue on eye l...

Page 8: ...eyepiece differ Inadequate illumination Lamp does not light Inadequate brightness Lamp blows out immediately Lamp flickers Power supply not plugged in Lamp not installed Lamp burnt out Specified lamp...

Page 9: ...tage is fashioned incorrectly Make sure cover glass is facing up Use a standard cover glass with a thickness of 0 13 0 17mm Turn the nosepiece into the right position until you hear it click into plac...

Page 10: ...t use any solvents other than absolute alcohol as they may damage lens bonding cement 3 Do not use petroleum benzene when cleaning components such as filters or lenses 4 Absolute alcohol and petroleum...

Reviews: