1.0 Introduction
The SpaceLogger.D10 cycle data logger
connects to the RS232 serial port on
sterilisers, washer disinfectors, ultrasonic
cleaners and other decontamination
devices. The logger captures the data
output by the machine and stores it on an
SD card. The SpaceLogger.D10 may be
set to provide a local alert to the operator of
a detected pass or failed cycle.
The cycle information can then be imported
into the SpaceLogger Cycle Manager
Software on a PC where the records may
be viewed, printed and archived.
2.0 Supplied Items
1. SpaceLogger.D10
2. Mains power adaptor
3. SD card
4. USB SD card reader
5. USB extension cable (for card reader)
6. SpaceLogger Cycle Manager Software
CD
7. Fasteners
3.0 Set Up
3.1 Connecting to the Decontamination
Device:
1. Connect the SpaceLogger.D10 to the
machine using the RS232 cable to plug into
the serial port.
2. Connect the mains power adaptor (to the
socket of the back of the logger) and plug
into a mains socket (100-240V). Switch on
power.
NOTE
– The SpaceLogger.D10 must be
configured for a specific machine before
recording data.
NOTE
– The RS232 cable supplied with the
SpaceLogger.D10 kit is specifically for use
with a particular machine and will not be
compatible with a different model. Please
check you have specified the correct
autoclave/washer model when ordering.
2
7
7.0 Specification
Physical
Dimensions: Width: 67 mm, Depth: 67 mm,
Height: 28 mm (excluding optional feet)
Weight: 75g
I/O Capability
Transmission standard: RS232 compatible,
selectable bits and parity.
Transmission speed: 9600 Baud (default)
or selectable from 110 to 115200 Baud
Data Storage
Data Storage Card: Removable SD card,
MMC or MMC mobile
Data Capacity: 2 GByte (max)
File System: FAT16 or FAT32 with 8.3 file
names, Sector size 512 Bytes
Visual Indicators
Card Status (front of logger):
Green LED: READY to record data
Red LED: DATA RECORDING to SD
card
Red LED on permanently (with audible
alarm): CARD ERROR
Cycle Pass/Fail Alert (keypad):
Green LED flash: Pass detected
Red LED flash: Fail detected
Audible Indicators
Audio Bleeper signals on: card insertion,
configuration, cycle pass/fail detection, card
error, off button depress
User Interface
Keypad with wipe clean Autotex®AM with
Microban® antimicrobial protection
Power
Power requirement: 7 to 30 Vdc
Current at 12Vdc: 10 mA typical
Supply input protection: Polarity reversal
protected and internal fuse
Connection: 1.3 mm centre pin DC
connector
Environmental
Temperature Range: Operating: -25 ºC to
+70 ºC, Storage: -40 ºC to +70 ºC
Enclosure protection: IP203
[Autotex® is a Registered Trademark of
MacDermid Autotype Ltd. Microban® is a
Registered Trademark of Microban Products
Company]
SpaceLogger.D10
with USB card
reader and SD
card