The following section provides installation instructions for the Summit SD-1150-WP dialysis wheelchair scale.

2.1 Scale

Standard installation of the scale consists of the following:

• Site preparation
• Unpacking the scale
• Assembly
• Electrical interface to the indicator
• Pit installation
• Load cell connections
• Powering the scale

2.2 Site Preparation

Consider the following when choosing a site for the scale:

• Select a site where overweight loads can maneuver easily without crossing the platform
• Avoid areas where damage could occur from side impacts of wheels or forklift tines
• Avoid areas where falling objects could cause shock damage
• Avoid areas where water may damage a scale not meant for a washdown environment
• The scale must be level within 1/4 inch of horizontal
• The interface cable between the scale and the indicator must be protected from crushing, cutting and moisture damage

The scale must not be loaded beyond capacity, even momentarily.
Choose a site where the floor is level to 1/4 inch to avoid excessive shimming. The floor may require 

modification if unable to select an area up to this standard.
If the chosen site has potential dangers to cable integrity, cable protection is required, such as running the cable 

in conduit.

2.3 Unpacking 

Remove all packing material and inspect the contents for damage possibly caused during shipment. Contact Rice Lake 

Weighing Systems and the shipper immediately if there is damage to the scale. The shipping container should contain the 

scale, pit frame, scale feet, this manual, the indicator, and a 10 ft length of load cell cable.
The Summit SD-1150-WP dialysis wheelchair scale has one threaded hole located in the center of the deck to allow for the 

installation of an eyebolt. An eyebolt hook and chains can be used to lift the scale. It is recommended to use a 1/2 inch-20NF 

eyebolt for lifting the scale.

Figure 2-1. Proper Lifting Technique

IMPORTANT

Summary of Contents for Summit SD-1150-WP

Page 1: ...Summit SD 1150 WP Dialysis Wheelchair Scale Software Version 11525 Operation Manual PN 197294 Rev A August 18 2022...

Page 2: ...respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserves the right to...

Page 3: ...ing or obtained by calling 715 234 9171 and asking for the training department Revision History This section tracks and describes the current and previous manual revisions for awareness of major updat...

Page 4: ...4 Assembly 4 2 5 Electrical Interface to the Indicator 4 2 5 1 Insert Batteries 5 2 5 2 Power Connection 5 2 6 Load Cell Connections 6 2 7 Pit Installation 7 2 8 Dimensions 7 3 0 Operation 8 3 1 Key D...

Page 5: ...nst harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance wit...

Page 6: ...rt the scale while someone is on the scale Do not allow minors children or inexperienced persons to operate this scale Do not use in the presence of flammable materials Do not use this product if any...

Page 7: ...the indicator must be protected from crushing cutting and moisture damage The scale must not be loaded beyond capacity even momentarily Choose a site where the floor is level to 1 4 inch to avoid exc...

Page 8: ...cess plate located on the side of the scale Use the following steps to wire the junction box 1 Remove the two 10 x 3 8 inch screws 2 Slide the junction box assembly out of the deck 3 Open and remove t...

Page 9: ...4 Battery Chamber 3 Put the cover in place and turn the thumbscrew clockwise to secure Remove the batteries prior to storing if the product is not going to be used for an extended period of time If t...

Page 10: ...to gain access to the load cell connection Figure 2 6 Remove Tilt Stand 2 Remove the four back retaining screws to remove the back cover to the indicator Figure 2 7 Remove Back Cover 3 Plug the end o...

Page 11: ...ale in a floor level pit and are for reference only the pit must be installed in poured concrete foundation according to standard construction practices Figure 2 9 Pit Frame Diagram A 1 24 slope is re...

Page 12: ...MI Pressing the BMI key enables access to the BMI Body Mass Index mode defaults when scale is turned on The patient is gets on the scale weight stabilizes and press the BMI key The display then asks f...

Page 13: ...y the patient 1 Press to turn on the scale 0 0 appears on the display along with the ZERO annunciator 2 Place additional item s on the scale 3 Press until the display returns to 0 0 and NET annunciato...

Page 14: ...adjust the height value 5 Press 6 The BMI value and BMI annunciator are shown on the display Press to return to the Weighing mode and the BMI function will be turned off 3 6 2 KG Mode 1 Ensure that t...

Page 15: ...high E07 AD too low E10 Overload Scale has been overloaded Remove load from scale E4L BAT Battery low but still usable one bar left on indicator display E4U Battery low and unstable no bars left on in...

Page 16: ...ess and hold for at least three seconds or until the scale emits two quick beeps If the scale does not beep after five seconds release as the weight was either in motion or out of range If displaying...

Page 17: ...the use of the USB driver Driver protections may need to be temporarily disabled on Windows 10 or later computers to allow for the installation of the USB driver 1 Connect the scale s indicator to a...

Page 18: ...the overall appearance of the entire scale for any obvious signs of damage Inspect the condition of the AC power adapter cord for cracking fraying or for broken or bent prongs 5 2 Cleaning Proper car...

Page 19: ...urs of continuous use Automatic power off can be configured Environmental Operating Temperature 50 F to 104 F 14 C to 40 C Storage Temperature 32 F to 158 F 0 C to 70 C Humidity 85 relative humidity D...

Page 20: ...Summit SD 1150 WP Wheelchair Scale 16 Visit our website www RiceLake com...

Page 21: ......

Page 22: ...54868 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems Specifications subject to change without notice www ricelake com PN...

Reviews: