SCT Weight Transmitter – 20 Series Technical Manual
46
Visit our website
At the time of writing, the setpoints, hysteresis values are saved to the RAM and will be lost upon the next
power-off; to store them permanently to the EEPROM so that they are maintained at power-on, the ‘99’ command
of the Command Register must be sent.
Weight (Span) Calibration Commands (With Test Weights)
The SCT20-MB calibration can be changed via Modbus. To perform this procedure, the system must be unloaded and the
weight value display reset to zero with the command 100 of the Command Register. A load must then be placed on the system
and the correct weight value must be sent to the registers 40037-40038. To save this value, send the control 101 from the
Command Register. If the operation is completed successfully, the two test weight registers are set to zero.
Analog Output Setting (Analog Models Only)
Write the weight in the registers 40045 (Weight value corresponding to the Full Scale of analog output H) and 40046 (Weight
value corresponding to the Full Scale of analog output L), or write the weight in the registers 40043 (Weight value
corresponding to the ZERO of the analog output H) and 40044 (Weight value corresponding to the ZERO of the analog output
L). After writing the value, send the command 99 from the Command Register to save it to EEPROM memory.
40013
PEAK WEIGHT L
—
R
40014
Divisions and Units of measure )
—
R
40015
Coefficient H
—
R
40016
Coefficient L
—
R
40017
Setpoint 1 H
Only after command ‘9’’ of the
COMMAND REGISTER
R/W
40018
Setpoint 1 L
R/W
40019
Setpoint 2 H
R/W
40020
Setpoint 2 L
R/W
40021
Setpoint 3 H
R/W
40022
Setpoint 3 L
R/W
40023
Hysteresis 1 H
R/W
40024
Hysteresis 1 L
R/W
40025
Hysteresis 2 H
R/W
—
40026
Hysteresis 2 L
R/W
—
40027
Hysteresis 3 H
R/W
—
40028
Hysteresis 3 L
R/W
—
40029
Inputs
—
R
40030
OutputS
NO
R/W
40037
Test weight for calibration H
Use with command ‘101’ of the
COMMAND REGISTER
R/W
40038
Test weight for calibration L
R/W
40043
Weight value corresponding to ZERO of
the analog output H
Ony after command ‘99’ of the Command
Register. (Analog Models Only)
R/W
40044
Weight value corresponding to ZERO of
the analog output L
R/W
40045
Weight value corresponding to Full Scale
of the analog output H
R/W
40046
Weight value corresponding to Full Scale
of the analog output L
R/W
REGISTER
DESCRIPTION
Saving to EEPROM
ACCESS
Table 3-33. Usable Registers (Modbus-RTU) (Continued)