1.2.9
Analog Output Menu
The ALGOUT menu is used only if the analog output option is installed. If the analog output option is installed,
configure all other indicator functions and calibrate the indicator (see Section 1.0) before configuring the analog
output. See Section 1.8 on page 46 for analog output calibration procedures.
Figure 1-13. Analog Output Menu
OFFSET ERRACT MIN
MAX TWZERO TWSPA
ALG OUT Menu
Parameter
Choices
Description
Level 2 submenus
SOURCE
GROSS
NET
Specifies the source tracked by the analog output.
OFFSET
0%
20%
Zero offset. Selects whether the analog output supplies voltage (0–10 V) or current (4–20 mA)
output. Select 0% for 0–10 V output; select 20% for 4–20 mA output.
ERRACT
FULLSC
HOLD
ZEROSC
Error action. Specifies how the analog output responds to system error conditions.
Possible values are:
FULLSC
Set to full value (10 V or 20 mA)
HOLD:
Hold current value
ZEROSC:
Set to zero value (0 V or 4 mA)
MIN
000000
number
Specifies the minimum weight value tracked by the analog output. Specify a weight value (in
primary units) in the range 0–999 999.
MINNEG
OFF
ON
Turn ON or OFF.
MAX
010000
number
Specifies the maximum weight value tracked by the analog output. Specify a weight value (in
primary units) in the range 0–999 999.
MAXNEG
OFF
ON
Turn ON or OFF.
TWZERO
—
Tweak zero. Adjust the analog output zero calibration. Use a multimeter to monitor the analog
output value. Press and hold or to adjust the output. Press to save the new value.
TWSPAN
—
Tweak span. Adjust the analog output span calibration. Use a multimeter to monitor the analog
output value. Press and hold or to adjust the output. Press to save the new value.
Table 1-9. Analog Output Menu Parameters