![Rice Lake Ishida IPC-WP 15 Service Manual Download Page 32](http://html1.mh-extra.com/html/rice-lake/ishida-ipc-wp-15/ishida-ipc-wp-15_service-manual_1454167032.webp)
29/32
IPC-WP Series Service Manual
3. Remove the 4 Sheet 3s.
Caution
When re-assembling the scale, use a new sheet
protect.
When affixing the new sheet protect, completely
remove the starch left on the bracket.
Otherwise, the water proofing property will be
deteriorated.
4. Remove the 4 Screws (M5x5)
The picture shows the position of the screws
(M5x5).
The two screws on the right side include a collar,
flat washer, and spring washer.
The other two screws on the left side include a
flat washer and spring washer, but no collar.
5. Cut the lock tie that is holding the cell cord
wires and grounding wires to the bottom of the
Platter bracket.
6. Pull out the Ferrite core that is inserted into the
cell cord.
Sheet 3
Collar
M5×15.SW.W
Ferrite core
Lock tie