![Rice Lake Ishida IPC-WP 15 Service Manual Download Page 19](http://html1.mh-extra.com/html/rice-lake/ishida-ipc-wp-15/ishida-ipc-wp-15_service-manual_1454167019.webp)
16/32
IPC-WP Series Service Manual
¡
Data after E2ROM is cleared
Mode
Data
Item
C1 1
JAPAN
C2
11
Scale No.= 3kg (2g/1g) Multi interval Decimal point indication= " 0"
C3
Zero point and span adjustment value and each approximate value
F 804
Measurement
condition
2.8.2 Error
No.
List
Error No.
Mode
Item
Err1
At power ON
E2PROM unsetting or garbled data
Err2
At power ON
Outside start range
Err3 Test
mode
Outside range where zero point can be adjusted (original A/D
1000 count or less)
Err4
Test mode
(Original A/D 25000 count or more)
Err5 Test
mode
Outside span adjustment possible range (original A/D 33300
count or more)
Err6
Normal mode Original A/D zero count or less
Summary of Contents for Ishida IPC-WP 15
Page 1: ......
Page 2: ......
Page 5: ......
Page 9: ...6 32 IPC WP Series Service Manual 1 4 Operation Panel ASIA Specification Oceania Specification...
Page 10: ...7 32 IPC WP Series Service Manual EU Specification USA Specification...
Page 11: ...8 32 IPC WP Series Service Manual 1 5 Outer Dimensions Unit mm...
Page 22: ...19 32 IPC WP Series Service Manual 3 1 2 Single Display...
Page 36: ......
Page 37: ......
Page 38: ...PN 86367 01 2012...