IQ plus 510/710 Analog Output Option

 

Enclosure Reassembly

 

Once cabling is complete, position the backplate over
the enclosure and reinstall the backplate screws. Use
the torque pattern shown in Figure 5 to prevent
distorting the backplate gasket. Torque screws to 10
in-lb (1.13 N-m).

 

Figure 5. IQ plus 510/710 Enclosure Backplate

 

Analog Output Calibration

 

The following calibration procedure requires a
multimeter to measure voltage or current output from
the analog output module. No test weights are
required for calibration.

 

NOTE:

 

  The analog output must be calibrated 

 

after

 

  the

indicator itself has been configured and calibrated.

 

Figure 6. IQ plus 710 Analog Output Menu

 

1. Enter setup mode and go to the ALGOUT

menu (see Figure 6).

Set OFFSET to 0% for 0–10 V output,
20% for 4–20 mA output

Set MIN to lowest weight value to be
tracked by the analog output

Set MAX to highest weight value to be
tracked by the analog output

To enter MIN and MAX values, use the   and

  keys to select the digit; use   and   to

increment or decrement the value.

2. Connect multimeter to analog output:

For voltage output, connect voltmeter
leads to pins 3 and 4

For current output, connect ammeter
leads to pins 1 and 2

3. Adjust zero calibration: Scroll to the

TWZERO parameter. Check voltage or
current reading on multimeter. Press and hold

 or   to adjust the zero value up or down.

4. Adjust span calibration: Scroll to the

TWSPAN parameter. Check voltage or
current reading on multimeter. Press and hold

 or   to adjust the span value up or down.

5. Final zero calibration: Return to the

TWZERO parameter and verify that the zero
calibration has not drifted. Press and hold 
or   to re-adjust the zero value as required.

6. Return to normal mode. Analog output

function can be verified using test weights.

S e t u p   s w i t c h   a c c e s s   s c r e w

Fillister head screws

1

To r q u e   b a c k p l a t e   s c r e w s

t o   1 0   i n - l b   ( 1 . 1 3   N - m )

2

3

4

5

6

7

8

9

10

11

16

12

13

14

15

XXXXXXX

XXXXXXX

ALGOUT

DIG IN

SETPTS

XXXXXXX

PROGRM

PFORMT

SERIAL

CALIBR

XXXXXXX

CONFIG

FORMAT

SOURCE

GROSS

0%

20%

OFFSET

ERRACT

FULLSC

HOLD

000000

number

MIN

10000

number

MAX

TWZERO

NET

TWSPAN

ZEROSC

XXXXXXX

VERS

71

number

15171

number

Summary of Contents for IQ plus 355

Page 1: ...on its standoffs in the location shown in Figure 1 and plug the module input into connector J5 on the IQ plus 355 board Connect output cable to the analog output module as shown in Table 1 then reassemble the enclosure as described on page 2 Table 1 Analog Output Module Pin Assignments Figure 1 Analog Output Module Installation on IQ plus 355 Board Caution Pin Signal 1 Current Out 2 Current Out 3 ...

Page 2: ...ment or decrement the value 2 Connect multimeter to analog output For voltage output connect voltmeter leads to pins 3 and 4 For current output connect ammeter leads to pins 1 and 2 3 Adjust zero calibration Scroll to the TWZERO parameter Check voltage or current reading on multimeter Press and hold or to adjust the zero value up or down 4 Adjust span calibration Scroll to the TWSPAN parameter Che...

Page 3: ...connector J5 on the IQ plus 510 710 board Connect output cable to the analog output module as shown in Table 2 then reassemble the enclosure as described on page 4 Table 2 Analog Output Module Pin Assignments Figure 4 Analog Output Module Installation on IQ plus 510 710 Board Caution Pin Signal 1 Current Out 2 Current Out 3 Voltage Out 4 Voltage Out ANALOG OUTPUT MODULE 100 76 75 51 50 26 25 1 U22...

Page 4: ...keys to select the digit use and to increment or decrement the value 2 Connect multimeter to analog output For voltage output connect voltmeter leads to pins 3 and 4 For current output connect ammeter leads to pins 1 and 2 3 Adjust zero calibration Scroll to the TWZERO parameter Check voltage or current reading on multimeter Press and hold or to adjust the zero value up or down 4 Adjust span calib...

Page 5: ...tput module on its standoffs in the location shown in Figure 7 and plug the module input into connector J5 on the IQ plus 350 board Connect output cable to the analog output module as shown in Table 3 then reassemble the enclosure as described on page 6 Table 3 Analog Output Module Pin Assignments Figure 7 Analog Output Module Installation and Wiring on IQ plus 350 Board Caution Pin Signal 1 Curre...

Page 6: ...se the and keys to select the digit use and to increment or decrement the value 2 Connect multimeter to analog output For voltage output connect voltmeter leads to pins 3 and 4 For current output connect ammeter leads to pins 1 and 2 3 Adjust zero calibration Scroll to the TWZERO parameter Check voltage or current reading on multimeter Press and hold or to adjust the zero value up or down 4 Adjust...

Reviews: