Troubleshooting
© Rice Lake Weighing Systems
●
All Rights Reserved
9
4.0
Troubleshooting
This section provides information regarding the troubleshooting of the Altralite Floor Scale.
4.1 Troubleshooting Guide
4.2 Periodic Maintenance
The space between the platform side and pit frame, and the surface beneath the platform must be periodically cleaned to
prevent debris build up. More frequent cleaning of these areas is necessary with scales mounted in pits.
Do not attempt to use scales with load cells that are not hermetically sealed in washdown applications.
Water damage is a common cause of failure in non-hermetically-sealed load cells. Use care with high pressure
steam wash downs for hermetically-sealed load cells. The steam may not damage the load cells, but the
elevated temperatures may cause incorrect readings until the unit cools to room temperature.
Issue
Possible Cause
Solution
System does not operate - no display Power disconnected
Check and reconnect
Indicator fuse blown
Replace fusel; Check for cause
Interface cable cut or disconnected
Repair
Signal leads incorrectly installed at indicator Install according to indicator installation manual
Display stays at zero
Indicator faulty
Service indicator
Load cell connections faulty
Check cable connections in junction box and at indicator
Erratic weights
Vibration near scale
Remove source of vibration or move scale
Platform not level to within 1/4''
Level scale by adjusting feet or shimming if necessary
Load cell or cable water damage
Replace
Debris under load cells or platform
Clean
Indicator faulty
Use simulator to test indicator for stability;
Service indicator
Consistently high or low weights
Indicator not properly adjusted to zero
Zero the indicator according to indicator manual
Platform binding
Obtain adequate clearance for free platform movement
Indicator not calibrated
Calibrate according to indicator manual and
Load cells are faulty
Test and replace load cells if necessary
Feet touching deck underside
Adjust feet downward to provide clearance
Table 4-1. Troubleshooting