Page 1: ...151806 Rev B Digital Wheelchair Scale w Seat 350 10 4 Software Revision 11439 Operation Manual...
Page 2: ......
Page 3: ...utton Keypad Print 11 5 2 USB Connection 11 6 0 Troubleshooting and Testing 14 7 0 Maintenance 15 7 1 Basic Maintenance 15 7 2 Cleaning 15 8 0 Digital Wheelchair Scale Specifications 16 For More Infor...
Page 4: ...ii Wheelchair Scale with Seat Operation Instructions Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com webinars...
Page 5: ...t of a moving patient The weight can be displayed in either pounds or kilograms and a tare weight can be entered The Scale Operation section on page 8 of this manual explains the scale operation and h...
Page 6: ...every individual who operates or works with this unit has read and understands the following safety information Please follow these instructions carefully Do not drop the scale or subject it to viole...
Page 7: ...e back Important Damage caused by improper packaging is not covered by the warranty 3 3 Setting Up Your Scale Use the following steps to set up the Digital Wheelchair Scale with seat 1 Locate the manu...
Page 8: ...counterclockwise two full turns This will ensure that there is enough clearance between the scale base and the floor 3 Gently set the scale base down to the floor There should be minimal clearance bet...
Page 9: ...ir scale by rotating the two handles This will secure the handrail mounted indicator Figure 3 3 Secure the Handrail by Tightening the Durable Hinges 3 The wheelchair scale with seat comes fully assemb...
Page 10: ...e retaining bracket is secure as shown in Figure 3 5 Caution The retaining bracket must click into place If it doesn t click it may not be secure and any weight applied to it may cause it to collapse...
Page 11: ...shown in Figure 3 7 Figure 3 7 Battery Chamber 3 Close the battery chamber If an external power supply or USB power supply is connected the battery flag on the display is turned off When using battery...
Page 12: ...to zero after three seconds It works only if the current weight is stable and zero up to 2 of full weight Hold Release Hold Release Hold Release The first press holds the most current weight value sh...
Page 13: ...Zero key will not work The only way to return to zero from here is to press the Hold Release key one more time Note Pressing the Hold key prior to a person getting on the scale will also work 4 3 Pres...
Page 14: ...MI and FT IN annunciators are lit on the display and a default value of 5 feet and 7 5 inches 5 07 5 is flashing 5 Use the Up Down arrow keys to adjust the height value and press the ENTER key to move...
Page 15: ...til the scale displays PRINT Note that if the scale does not beep after five seconds then release the button as the weight was either in motion or out of range If displaying weight and not BMI the sca...
Page 16: ...software aspx Select Medical Health Scales Software and Get Downloads Opening any product will show a USB Driver download Click on Download to open and download the driver to your computer 1 The graph...
Page 17: ...B cable is properly connected and unit is on 9 Another terminal type program such as Hyperterminal needs to be opened and connected through the USB driver to the indicator to be able to see the inform...
Page 18: ...the cables and mechanical connections If the problem persists replace the set of load cells Err 3 High saturation state high A D See Err 2 Err 6 Unstable weight Cannot calibrate Check the load cells...
Page 19: ...c power off can be configured Data Communications RS 232 with RJ 45 jack USB Connection Selectable baud rate default 9600 8 bits No parity 1 stop bit No handshaking Environmental Operating Temperature...
Page 20: ...ves are available 6 30 a m 6 30 p m Monday through Friday and 8 a m to 12 noon on Saturday CST Telephone Sales Technical Support 800 472 6703 Canada and Mexico Customers 800 321 6703 International 715...
Page 21: ...nonconformity actually exists and was not caused by accident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of all alleged non conform...
Page 22: ...18 Wheelchair Scale with Seat Operation Manual...
Page 23: ......
Page 24: ...e Lake WI 54868 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 88 2349171 www ricelake com www ricelake mx www ricelake eu www ricelake co in m ricelake com PN...