Dietary/Fitness Technical and Operating Instructions - Scale Configuration
13
4.4
Reset To Factory Defaults
The configuration parameters may be reset to factory defaults while in the configuration mode. To do so, enter
the configuration mode. Press the
REWEIGH
key once to advance to displaying the software version (ID <->
11305). Press and hold the
BMI
key until
=DEF=
is displayed. At this point, if you wish to perform the default
function, press the
REWEIGH
key and the scale will show
DONE
, then it will restart. If you do not want to perform
the default function, press the
ON/OFF
key.
After performing the default function, the scale will require re-calibration. In addition, the scale will be set
up to a capacity of 600 lbs. The capacity varies by model - refer to the serial label on your scale and
reconfigure the capacity (FULL) and display divisions (rOUnd) as necessary in the configuration menu.
This procedure also resets the Deep Programming Mode settings to factory defaults (see Section 2.1).
4.5
Deep Programming Mode
Before the scale can enter into the deep programming mode, the configuration enable jumper must be removed.
Use the following steps to access that jumper and get into deep programming mode.
1. Access to that jumper is gained by removing the back cover of the indicator. Remove the four back cover
retaining screws as shown on the left side of Figure 6?-?? on page 7. With the cover removed, the jumper
can be seen sticking through the hole in the rear housing (shown in Figure 6?-?? on page 7 - right side).
2. Remove that jumper to gain access to the deep programming modes.
3. Turn unit over and press and hold the
REWEIGH
key and the
On-Off/Zero
key simultaneously to turn the
scale on.
4. Continue to hold the
REWEIGH
key until
IDENT
appears on the display.
5. With
IDENT
appearing on the display, press and hold the
BMI/Tare
key until the
HP-x
appears on the
display which is the deep programming mode.
6. To change from one parameter to the next, you must press once on the
REWEIGH
key.
7. To change the value of the parameter, use the
Kg-Lb
key.
8. From the
SAVE
phase: press the
REWEIGH
key to exit and save the modified date.
DONE
appears for one or
two seconds followed by
Start
and the display enters into weighing mode and is ready to start the
weighing process.
9. Replace the programming/calibration jumper.
10. Replace the back cover of the indicator and secure with screws.
11. Various parameters can now be set up while in programming mode.
NOTE:
A display of Con En indicates that the jumper is not in place. Put jumper on both pins to return the scale to
normal weigh mode.