Section 4
Saturn
™
5 Laser Systems Product Overview
4
16
Research Instruments Ltd
Control unit
220mmx180mmx34mm (8.6”x7.1”x1.3”)
Weight
1.2Kg
Dimensions
Laser Specifications
Laser Specifications
Ablation Laser
Pilot Laser
Output wavelength
1480nm
650nm
Power output from patch lead
400mW
180µW (max)
Maximum pulse width
2.0ms
None
Laser Safety Classification*
Class 1
Class 1
* IEC 60825-1:2014, US 21CFR 1040.10
Electrical Specifications
Mains Adaptor for Control Unit and Motor Module
Input
100-240VAC, 50-60Hz, 1.0 - 0.5A
Output Power (maximum)
40W
Voltage
12VDC
Current (maximum)
3.3A
Earth Connection
Mandatory 3-pin plug for earthing (grounding)
Range of Movement
Approx. 370μm (40x), 740μm (20x)
Accuracy
1μm within calibration area, 3μm outside (see page 33)
Saturn Active Specifications
Manufacturer
Actual Magnification
Olympus
40x, 20x
Nikon (CFI60)
36x, 18x
Leica
36x, 18x
Zeiss
44x, 22x
Magnification
40x
20x*
N.A.
0.49
0.25
Working distance
2.5mm
2.5mm
Parfocal distance
45mm (adaptor supplied for Nikon CFI60 systems)
Objective Specifications
*
Not cleared for use in USA market.
40x or 20x is a nominal figure. Actual magnification will vary slightly when fitted to different microscopes.
Temperature
10°C (50°F) to 42°C (108°F)
Humidity
15% to 85% RH (Non-condensing)
Pressure
Pressure Range: 70kPa. to 108kPa
Operating Conditions
Temperature
-40°C (40°F) to 60°C (140°F)
Humidity
15% to 85% RH (Non-condensing)
Pressure
Pressure Range: 70kPa. to 108kPa.
Storage/Transport Conditions