![RHEONIK RHE28 Installation & Start?Up Manual Download Page 34](http://html.mh-extra.com/html/rheonik/rhe28/rhe28_installation-and-start-up-manual_1452362034.webp)
Appendix A: Safety Instructions – Ver. 1.00
34 / 43
A.10
Technical Data
The following limits are valid for safe operation:
Electrical data:
Refer to chapter A.3 and to the information on the letter plate
Operating temperatures:
-20°C (-40°C) to +60°C (version dependent)
Humidity:
10 to 90% relative humidity, non-condensing
Altitude:
up to 3000m, higher maximum altitude upon request
For the complete technical data refer to appendix B “Technical Data”.
A.11
Compliance
The RHE28 certified according to ATEX / IECEx comply with the following standards for hazardous
areas:
IEC 60079-0
IEC 60079-15
IEC 60079-11
For the list of standards relevant for a certain unit and for the release date refer to the certificate.
The RHE21 certified by CSA comply with the following standards for hazardous areas:
C22.2 No 61010-1: 2012
UL 61010-1 3rd Ed.
C22.2 No 30-M1986 (R2012)
UL 1203, 5
th
Ed.: 2013
C22.2 No 60079-0, 6
th
Ed.
UL 60079-0, 6th Ed.
C22.2 No 60079-11, 6
th
Ed.
UL 60079-11, 6
th
Ed.
C22.2 No 60079-15, 4
th
Ed.
UL 60079-15 4
th
Ed.
C22.2 No 213 / ANSI/ISA 12.12.01: 2017
C22.2 No 60529 Ed. 2.2.
ANSI/ISA 60529 Ed. 2.2
C22.2 No. 94.2-07 (R2012)
UL 50E: 2015
For the list of standards relevant for a certain unit and for the release date refer to the certificate.