Testing and Care
Page 22 of 28
Baby Check Transmitters User Guide
0510-1124-I
4. While holding the button in, the indicator
light flashes and a tone sounds once
per second.
5. Wait for at least 3 flashes of the
indicator light and 3 tones from the
transmitter tester to verify that the
clearing device is functioning correctly.
6. If the device fails this test, check the
battery in the tester. The transmitter
tester requires a 9V battery.
7. Retest the device. If it fails a second
time, DO NOT USE THE
TRANSMITTER CLEARING DEVICE
and contact RF Technologies Technical
Support.
8. Your facility must keep records of test
and transmitter inspection.