Operation and Maintenance Manual – Steam sterilizer mod. Matika
Rev. 05 of 01-12-2006 - C:\Documenti\Manuali di istruzione\Matika GB-R05.doc
Page 12 of 31
5.2
Installation and preliminary inspections
The installer is obliged to make sure that the place in which the sterilizer is installed
is fit for that purpose (CEI 62/4).
The machine is sent packed on a wooden pallet. Make sure that the packing has not been damaged and
that everything is in a perfect condition. Handle the box by means of its handles and place it on a
surface in the immediate vicinity of the place in which the machine is to be positioned (the weight of
the model is given in chap. 2, in the technical specifications).
Remove the machine from its wrapping by pulling it out of the box vertically, using 2 hooks (available
on the market) inserted into the two holes at the sides of the casing. Remove the polystyrene filler balls
from around the machine. Make sure that none of the parts has been damaged and that the machine is
complete with all the accessories indicated in the consignment note. Immediately contact our reseller if
discrepancies are discovered.
All the packing materials must be kept for guarantee purposes. Machines returned without their original
packing will not be accepted.
Install the autoclave in a well ventilated place, well away from sinks, heat sources, grinders and from
all other machines that create vibrations and/or dust.
Place the autoclave delicately on to a perfectly flat surface. Take care to leave a space of about 5-10 cm
all around the machine to allow for aeration.
Autoclave mod. MATIKA must be installed in a well lighted place equipped with an electrical system
that complies with the current standards in force.
Remember that MATIKA cannot be built into other furniture as it is filled with water from the top.
Fit the electric plug into a socket
WITH A CAPACITY OF AT LEAST 16 Ampere AND
EQUIPPED WITH AN EARTH CONNECTION
. Make sure that the mains voltage rating
corresponds to the value indicated on the label at the rear of the machine (230 Volts).
MATIKA
must only be filled with
DISTILLED WATER
to ensure that the electrical and air-operated
devices operate efficiently and are long lasting.
Carry out a cycle of dry-run tests using one indicator (type TERMALOG or VAPOR LINE
EXTENDER).
Keep the receipt which should be attached to the installation form and sent off along with the warranty.
Never connect the plug of the machine to a reduction.
Do not install the machine in places at altitudes of more than 1200 m.
above sea level.
If this is the case, contact the manufacturer for a special version of the
machine.
Use a bubble level to make sure that the surface on which the appliance is
to be positioned is perfectly horizontal.
Fill out and send off the warranty form.