Operation and Maintenance Manual – Steam sterilizers mod. Full Clave
Rev. 01 of 01-12-2006 - C:\Documenti\Manuali di istruzione\FULL CLAVE GB-R01.doc
Page 13 of 34
6 .
G E N E R A L
I N S T R U C T I O N S
F O R
U S E
6.1 Directions for use
OPEN THE DOOR BY MEANS OF THE HANDLE (12).
-
Remove the tray holder package from the chamber.
-
Turn on the main switch (17).
-
Pull the water fill tank (13) fully out (about 5 cm). Unscrew the black plug and pour in distilled
water until the maximum level indicator light (7) comes on and the buzzer operates. Screw the plug
back on and set the tank back in its original position.
-
Press the ON/OFF button (11) to activate the steam generator. Indicator light (5) will come on.
-
When the generator has reached the correct temperature, the number (from 1 to 11) of the last cycle
used will appear on the display (1) along with the relative description.
-
To choose a different cycle, press the SELECT button (9) to select the required program (from 1 to
8) and confirm with (10). Program number 10 is the Bowie & Dick Test while program number 11
is the Vacuum Test.
-
If a customized (“special”) cycle is required, select number 9, confirm by pressing (10) and comply
with what happens after having shut the door with the handle (12):
-
Press buttons (3) and (2) for about 15 seconds at the same time until the values on the display (1)
begin to flash.
-
Now edit the data entered with keys (3) and (2). Confirm with button (10) once the required value
has been reached.
-
The following cycle values can be modified in succession: Sterilizing temperature, Sterilizing time,
Drying time, number of vacuum peaks (from 0 to 3). It is advisable to consult the table on page 16.
-
Once the modifications have been made, fit the tray holder into the chamber with the items to be
sterilized, shut the door with the handle (12) and make sure that indicator light (6) comes on to
show that the door has been closed correctly.
-
Press button (10) to start the cycle.
-
The buzzer will operate and a message will appear on the display (1) to indicate that the cycle has
terminated.
-
Lack of clean water will be signalled by indicator light (8) and the buzzer.