Technical Data
12
3.5 Dimensions and Weight
Height:
288 mm
Height with test sieve (50 mm):
~ 382 mm
Width:
460 mm
Width with cyclone separator:
~ 665 mm
Depth:
302 mm
Depth with manual vacuum regulation:
~ 392 mm
Depth with automatic vacuum regulation: ~ 418 mm
Weight:
~ 16 kg
3.6 Required Floor Space
CAUTION
C6.0047
Falling down of the device
Incorrect positioning or insufficient working space
Due to its weight, the device can inflict personal injury if it falls down.
•
The device must only be operated on a sufficiently large, strong and
stable workplace.
•
All feet of the device must be positioned securely.
Width of the base: 480 mm
680 mm (with cyclone separator)
Depth of the base: 460 mm
No safety clearances required
Location requirements:
The device must be placed on a vibration-free, plane, stable and free surface to avoid
transmission of vibrations. A level base ensures the uniform distribution of the sample over the
sieve mesh fabric, as well as the stability of the device.
3.7 Receptacle Volume
The maximum receptacle volume (the maximum feed quantity) depends on various factors such
as aperture size of the test sieve, maximum grain size and width of distribution of the sample
material.
Examples for the maximum feed quantity according to DIN 66165 for test sieves of 200 mm in
diameter are listed in the following table:
Mesh size
Max. feed quantity
Max. permitted oversize material
according to DIN 66165
25 µm
14 cm
3
7 cm
3
45 µm
20 cm
3
10 cm
3
63 µm
26 cm
3
13 cm
3
125 µm
38 cm
3
19 cm
3
250 µm
58 cm
3
29 cm
3
500 µm
88 cm
3
44 cm
3
1 mm
126 cm
3
63 cm
3
Summary of Contents for AS 200 jet
Page 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...
Page 5: ...Notes on the Manual 5...
Page 25: ...First Commissioning 25 Fig 10 Example of a label on the sieve lid...
Page 42: ...Controlling the Device 42 Fig 21 Submenu SERVICE H7 1...
Page 61: ...61...
Page 62: ...62...
Page 63: ...63...
Page 64: ...Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...