Four Pump Linear Gradient Control

 

•  Up to four pumps can be controlled and monitored in a time based linear gradient.  

 

•  – Advanced       ----  Quad Pump Control

 

 

•  Virtual Pump will be shown (at the top of the form) only for the number of pumps detected on COM ports.

 

•  Gradient setups can be saved to text files and reloaded when needed. 

 

•  Gradient percentages will be calculated as they are entered into the chart.

 

•  100% will be set for the first solvent; enter the needed percentage starting at the second solvent. 

 

•  A “Graphical Gradient Display” tab is located on Gradient Table.  This will plot the programmed gradient

 

  for the user to verify the entered data.

Selection button will 

open a setting entry.

Specific pump 

options detected.

Graphing controls 

for user.

Graph may be saved or 

printed for future display.

Timer control will be 

available if a time 

value is entered.

Setup windows: 

enter settings and 

use the send button. 

Green indicates the 

row in use.

The delay or transition time

of the relay switching may be

modified if needed.

Summary of Contents for 26408

Page 1: ...www restek com 1 Pack In A Box Dual Piston Pump Operator s Manual cat 26408 Chromatography Products www restek com 1 800 356 1688 1 814 353 1300...

Page 2: ...2 4 4 Solvents With Harmful Effects 2 5 Instrument Installation 2 5 1 Mobile Phase Reservoirs 2 5 2 Inlet Tubing and Filters 2 5 3 Outlet Tubing 2 5 4 Priming the Pump 2 5 5 Long Term Pressure Calibr...

Page 3: ...eals 5 4 Check Valve Cleaning 5 5 Cleaning the Pump 5 6 Cleaning the Cabinet 5 7 Lubrication 5 8 Fuse Replacement 5 9 Battery Replacement 6 Quick Guide to Problem Solving 7 List of Replacement Parts A...

Page 4: ...without pulse damper 0 to 6000 psi with pulse damper Pressure Accuracy 1 of full scale pressure Pressure Zero Offset 0 2 psi Pressure Control Accuracy 10 or 100 psi whichever is larger Dimensions 5 5...

Page 5: ...r bubbles are formed during the inlet stroke The key to preventing cavitation is to reduce inlet restrictions The most common causes of inlet restrictions are crimped inlet lines and plugged inlet fil...

Page 6: ...acy This note applies if your pump is equipped with an electronic pressure transducer The transducer has been zeroed and calibrated at the factory Over the life of the pump some drift may occur For ex...

Page 7: ...actual flow rate The actual flow rate is displayed using the MODE key when the mL min indicator is blinking 3 1 3 1 Digital Display The 4 digit display shows the pump s maximum flow rate mL min pressu...

Page 8: ...s percentage use the up arrow and down arrow buttons When you have selected the desired value for ramp speed adjustment press the RUN button to return to the normal operation of the pump The 0 or 1 is...

Page 9: ...ll not have to be removed to read the EPROM label Serial Port Loop Back Test Mode If an external device will not communicate to the pump via the serial port the serial port loop back test can be used...

Page 10: ...plugged if the pres sure exceeds the maximum pressure rating of the pump or if the mechanism jams In the event a motor stall occurs the electrical current being supplied to the motor is turned off and...

Page 11: ...ure into the housing and tighten 5 2 Changing Pump Heads 5 2 1 Removing the Pump Head 1 Turn OFF the power to the dual head pump 2 Unplug the power cord 3 Remove the inlet line and filter from the mob...

Page 12: ...ng any other tool will scratch the finish 3 Repeat the procedure for the low pressure seal in the flush housing 4 Inspect and if necessary clean the pump head as described in Section 5 2 2 5 2 3 2 Cle...

Page 13: ...2 mL SS 2000 psi 3 mL min 3000 4000 psi 3 4mL min 5 4 Check Valve Cleaning Many check valve problems are the result of small particles interfering with the operation of the check valve As a result mos...

Page 14: ...check the fuse located on the motor power circuit board Replace it with a 5 A slow blow fuse 5 9 Battery Replacement The battery provides power for the memory that holds the current pump configuratio...

Page 15: ...4 Consider changing to self flushing pump 5 Excess lubricant on piston head if using buffers carrier 5 Clean excess lubricant and dirt from piston carrier See Section 5 8 Blue dye in mobile phase Puls...

Page 16: ...Transducer Tee Assembly 880139 Dual Head SMT Board Set pump serial 20 000 880112 Dual Head Board Set pump serial 20 000 880122 Front Panel Assembly 880904 Dual Head Overlay Part number for single pie...

Page 17: ...racterized by a unique two letter command code and only one command can be issued per line Case is not important that is the command codes PR Pr pR and pr are all equivalent Response strings sent by t...

Page 18: ...the minimum value is 0 The value must be expressed as four digits i e for 100 psi xxxx 0100 SF OK Puts the pump in fault mode Turns on the FAULT LED and stops the pump immediately RF OK x y z Reads t...

Page 19: ...providing the proper run stop logic level flow rate control frequency or flow rate control voltage can be used as a remote con trolling device for pump operation via this connection The terminal boar...

Page 20: ...contact an open collector output an open drain output or any output with a high logic level output of 3 8 to 6 0 volts and a low logic level output of 0 0 to 0 5 volts A switch contact or a relay cont...

Page 21: ...which drives the VOLTAGE IN and VOLTAGE COM connections must be isolated from the safety ground to prevent a ground loop current If the pump s displayed flow rate jumps up and down erratically suspect...

Page 22: ...with firmware identification Information screens are updated automatically however refresh buttons are available for pump information Basic operations and basic flow rate buttons are available Flow ra...

Page 23: ...e above When time events are programmed the Timer control will appear in the top middle of the screen Up to 3 timed triggers of the pumps relay output may be set Note Pump firmware must be compatible...

Page 24: ...0 will be set for the first solvent enter the needed percentage starting at the second solvent A Graphical Gradient Display tab is located on Gradient Table This will plot the programmed gradient for...

Page 25: ...bsolute accuracy of the polling interval is not guaranteed this is a Windows limitation The graph will start when the Poll for Flow and Pressure button is pressed To log the data click the Log Data bu...

Page 26: ...gure Assign Com Ports PC connection to the pump A wiring diagram is available under Configure RS232 Connection A timer is available under Testing Timer Basic up down functions are available The PC s s...

Page 27: ...m date of shipment In the event of such a defect Restek will repair or replace the product or necessary parts therein at its discretion and such repair or replacement shall be the sole remedy of this...

Page 28: ...fax 1 814 353 1309 www restek com Restek France phone 33 0 1 60 78 32 10 fax 33 0 1 60 78 70 90 e mail restek restekfrance fr Restek GmbH phone 49 0 6172 2797 0 fax 49 0 6172 2797 77 e mail info rest...

Reviews: