S E T U P F U N C T I O N
During changing the different program version IC or some abnormal voltages and
operations, you need to SETUP the DRO.
Procedure:
1). Turn off the DRO.
2). Power on the DRO once again, when the messag
e "VER.9GA" showing
in the MESSAGE WINDOW, please Enter key immediately then the DRO
starts entering into the SETUP function.
The SETUP procedure is written in a menu mode which enables you to scroll through
the top level options and enter, configure and exit the sub-functions as they arise.
When the DRO displays
some messages over,
it will into the
"SETUP" function.
M E S S A G E W I N D O W
A X I S K E Y
M E S S A G E W I N D O W
U P / D O W N K E Y S
E N T E R K E Y
The top level menu headers in order are as follows:
DIRECTIN
LIN COMP
QUIT
specifies the direction of count for each axis
permits linear error compensation to be input
exits the SETUP function to proceed to normal working
NL Error
FILTER
permits non linear error compensation to be input (point compensation)
permits the filter value to be set