background image

- 6 -

X

X

ref

R

esson

Technologies

Summary of Contents for RD-14G

Page 1: ...he People that Always Committed to Quality Technology Innovation Technologies RD 14G Resson Technologies ent I O On Off Resson Technologies Co Ltd Digital Readout System Operation Manual Grinder RD 14...

Page 2: ...ductor can render the equipment potentially dangerous Any intentional break is not permissible Please install the Linear Encoder finish after that switch on the Digital Readout Unit If switches on Dig...

Page 3: ...Display function 8 LED Response speed 60m 198 6feet min Quantizing error 1 count Power source AC 0V 0V 50 60Hz 20VA 10 23 Temperature fange Service 0 40 Storage 20 70 Linear Encoter Scales Electrical...

Page 4: ...Resson Technologies Content 1 Basic Functions 1 2 REF Datum Memory 6 3 Non Linear error Compensation 10 4 RD 14G Setup Function 17...

Page 5: ...1 Resson Technologies Basic Functions Basic Functions RD 14G Resson Technologies ent I O On Off...

Page 6: ...Set the current position for that axis to an entered Dimension Purpose 2 RD 14G Resson Technologies ent I O On Off 112 285 50 110 ABS RD 14G Resson Technologies ent I O On Off ABS 0 000 50 110 RD 14G...

Page 7: ...rk piece datum INC work piece zero position is still retained in coordinate by the ABS RD 14G Operator can then toggle between absolute and incremental ABS INC coordinates without losing the work piec...

Page 8: ...2 Locate the edge finder at the opposite end of the work piece Step 3 Then half the display coordinate using center find function as per follows Now the X Axis zero point 0 000 is located right at th...

Page 9: ...n filtering mode Example 1 The tum vibration filter on Example 2 Vibration filter currently to turn it on off Purpose The readout offer 199 subdatum memory function it is offered as a supplement of AB...

Page 10: ...6 X X ref Resson Technologies...

Page 11: ...he DRO system is switched off Therefore we simply need to store the distance between the point and the work piece datum zero ref position in DRO s memory Then in case of the power failure or the RD 14...

Page 12: ...e switches on the RD 14G step 3 move the machine reader head across the length of the transducer until digits display in RD 14G start to count step 1 To enter the ref function select the find mark FIN...

Page 13: ...the transducer select the RECALL 0 recall work zero step 2 select the axis of which work datum zero position needed to be recovered enter into function ref RECALL 0 recall work zero FIND REF find mark...

Page 14: ...nologies Non Linear error compensation 0 0 0 0 0 0 NonLinearerrorcompensation error start point of error curve Compensation saeps 10 Curve Line Position zero 1 2 3 4 CP START CP PITCH CP STEP REF 5 6...

Page 15: ...eadout s CPU then compensate the readout readings according to the error table that built in during the set up process The compensation always started vital important to have the CP START position loc...

Page 16: ...l of 300mm as our measurement standard The step pitch of the step gage is 25mm Our maximum machine travel is 265mm Therefore CP PITCH 25mm CP STEP 265 25 10 6 steps since steps must be in integer then...

Page 17: ...CP START position 115 875 swap to INC coordinate zero at the CP START position in INC coordinate 3 START measure the error to build up a error curve To mark the error measurement more easy swap to IN...

Page 18: ...hen the readout display the software version number press ent once to enter the SET UP function 25 000 25 008 50 000 50 004 75 000 75 017 100 000 99 995 150 000 150 012 200 000 199 988 125 000 125 002...

Page 19: ...rror function enter the CP START next step enter the CP STEP next step Press to confirm the choice MEAS VAL is Measured Value next step RD 14G Resson Technologies ent I O On Off NL Error ent ent RD 14...

Page 20: ...up procedure Therefore after entered the error curve data the display must be switched off at least one time after enter the MEAS VAL mode press up or down button to select the step point value X axis...

Page 21: ...gies ent I O On Off SETUP DIRECTIN LIN COMP QUIT specifies the direction of count for each axis permits linear error compensation to be input exits the SETUP function to proceed to normal working NL E...

Page 22: ...O once again when VER moving in display window please press 8 key then DRO enters into the RESET function power on the DRO then DRO starts the self test function when VER 9GA appears in display window...

Page 23: ...he parameter has reset The reset of parameter has completed then start to enter into the test procedure of display light Turn off the DRO after complete the reset then power on the DRO once again RD 1...

Page 24: ...which enables you to scroll through the top level options and enter configure and exit the sub functions as they arise When the DRO displays some messages over it will into the SETUP function MESSAGE...

Page 25: ...or Y Axis 4 Press key to make your setting then press key to select the LIN COMP linear compensation function ent RD 14G Resson Technologies ent I O On Off DIRECTN RD 14G Resson Technologies ent I O O...

Page 26: ...ue of X Axis 30 through the AXIS key make a same procedure for Y Axis ent 5 Press key to make you setting then press key to select the NL ERROR non linear error compensation function enter the CP STAR...

Page 27: ...e X axis measured value Z axis measured value after all measured value entered into the readout press ent to exit the NL Error function RD 14G Resson Technologies ent I O On Off 0 000 0 000 CP STEP en...

Page 28: ...will exit the SETUP function and return to the ABS state By pressing the ENTER key the DRO exits the SETUP program and is ready for machining operations If use the error compensation function you mus...