22
c
leanIng
and
m
aIntenance
Cleaning
Warning: Disconnect the power cord from the electrical outlet before cleaning the housing of the UltraFill Station.
Warning: Do not use any oil, grease, or petroleum-based products on or near this unit.
The outer housing of the UltraFill Station may be cleaned with a mild household detergent and a damp cloth or
sponge. Avoid using large quantities of water to clean the unit and do not use any petroleum-based solvents or
cleaning agents.
Wipe the device completely dry after you have finished cleaning it.
The Philips Respironics UltraFill Oxygen Cylinders should be kept clean and free from moisture and dust. Clean the
cylinders as needed by wiping with a dry, lint-free cloth. Do not clean with a solvent based cleaning solution. Avoid
getting fluids or debris such as sand or dirt inside the oxygen cylinder.
Maintenance
The UltraFill device requires no periodic maintenance during or in between patient use. There is a bacteria filter
located inside the UltraFill, which can be changed at the provider’s discretion.
The UltraFill cylinders utilize an Aluminum alloy. They should undergo hydrostatic testing every 5 years. Cylinders that
have not been tested for over 5 years should not be used.
The “Cylinder Specification” section of this manual outlines the use and storage environmental conditions. Care
should be taken to ensure this is followed.
t
ranSportIng
tHe
u
ltra
F
Ill
S
yStem
Warning: to prevent an accidental tip over, do not attempt to move the UltraFill
when it is connected to a cylinder.
During operation, the device is should be upright with the all four wheels
contacting the floor.
Use the handles on either side of the UltraFill device to lift it (see
illustration). The device can be transported in any orientation. Verify the
protective cap is covering the oxygen cylinder connector before transport.