Amara View minimal contact full-face mask

Instructions for Use

Intended Use

This mask is intended to provide an interface for application of CPAP or bi-level therapy 

to patients.  The mask is for single patient use in the home or multi-patient use in the 

hospital/institutional environment.  The mask is to be used on patients (>66lbs/30kg) for 

whom CPAP and bi-level therapy has been prescribed.

Caution: 

U.S. Federal law restricts this device to sale by or on the order of a 

physician. 

Note

:

 

An exhalation port is built into this mask so a separate exhalation port is not required.

 

This mask is not made with natural rubber latex or DEHP.

Symbols

Warning or Caution

Note

Tip

Consult Instructions for 

Use

Not Made with Natural 

Rubber Latex

X1

 System One Resistance Control

 Warnings

• 

This mask is not suitable for providing life support ventilation.

• 

This mask is designed for use with CPAP or bi-level systems recommended by your 

health care professional or respiratory therapist. Do not wear this mask unless the CPAP 

or bi-level system is turned on and operating properly. Do not block or try to seal the 

exhalation port. 

Explanation of the Warning: 

CPAP systems are intended to be used 

with special masks with connectors which have vent holes to allow continuous flow 

of air out of the mask. When the CPAP machine is turned on and functioning properly, 

new air from the CPAP machine flushes the exhaled air out through the attached 

mask exhalation port. However, when the CPAP machine is not operating, enough 

fresh air will not be provided through the mask, and exhaled air may be rebreathed. 

Rebreathing of exhaled air for longer than several minutes can in some circumstances 

lead to suffocation. This warning applies to most models of CPAP systems.

Summary of Contents for Amara View

Page 1: ...rnings This mask is not suitable for providing life support ventilation This mask is designed for use with CPAP or bi level systems recommended by your health care professional or respiratory therapis...

Page 2: ...continue use and contact your healthcare professional Consult a physician if any of the following symptoms occur Unusual chest discomfort shortness of breath stomach distension belching severe headach...

Page 3: ...Anti Asphyxia Valve The anti asphyxia valve consists of an air inlet and a flapper b With the airflow turned off verify that the elbow with the flapper is lying flat so that room air can flow in and...

Page 4: ...ekly The headgear does not need to be removed for daily cleaning of the mask 1 Hand wash mask and headgear in warm water with liquid dishwashing detergent 2 Rinse thoroughly Air dry completely before...

Page 5: ...er your head e 2 Mask cushion Place your nose over the cushion nostril opening Push the mask cushion up Note Do not place your nose in the cushion nostril opening 3 Headgear clips Press into place to...

Page 6: ...e Some air leaking is normal Comfort Tips The most common mistake is overtightening the headgear The headgear should fit loose and comfortable If your skin bulges around the mask or if you see red mar...

Page 7: ...he headgear clips onto the mask frame 3 Mask Cushion Grasp both sides and push it onto the mask frame 4 Quick Release Tube Push onto the elbow until the quick release tabs click into place Philips Res...

Page 8: ...curve shown below is an approximation of expected performance Exact measurements may vary Pressure Flow Curve Flow Rate SLPM 0 0 10 0 20 0 30 0 40 0 50 0 60 0 70 0 0 5 10 15 20 25 30 Pressure cm H2 O...

Page 9: ...er damage caused by accident misuse abuse negligence alteration failure to use or maintain the Product under conditions of normal use and in accordance with the terms of the product literature and oth...

Page 10: ...r rights under this limited warranty contact your local authorized Respironics Inc dealer or Respironics Inc at 1001 Murry Ridge Lane Murrysville Pennsylvania 15668 1 800 345 6443 USA and Canada only...

Page 11: ...11...

Page 12: ...12 1112274 R05 AD 8 26 2016 2016 Koninklijke Philips N V All rights reserved...