6
Using the Stellar device for the first time
When using the Stellar device for the first time, ensure that all components are in working condition and
perform a functional test (refer to Performing a functional test (see page 9)). Alarms should also be
checked by the physician when setting up a new patient (see Clinical Guide).
Working with other optional accessories
Attaching the pulse oximeter
Use of a pulse oximeter may be recommended by your clinician.
Both the SpO
2
and pulse rate are logged at a sample rate of 1 Hz.
The optional SpO
2
sensor is calibrated for the display of functional oxygen saturation.
The Disconnected XPOD alarm will activate if the pulse oximeter is connected during therapy and an
interruption to its data transfer occurs for more than 10 seconds.
For part numbers of compatible oximeter accessories for this device, see the ventilation accessories on
www.resmed.com/accessories. For information on how to use these accessories, refer to the user guide
that comes with these accessories.
WARNING
•
Pulse oximeter sensors must not be used with excessive pressure for prolonged periods as this
can cause patient pressure injury.
•
The pulse oximeter sensor and cable needs to be verified for compatibility with Stellar,
otherwise, patient injury can result.
Contraindication
The pulse oximeter does not meet the defibrillation-proof requirement as per IEC 60601-1: 1990, clause
17.h.
Refer to illustration G.
1.
Connect the plug of the finger pulse sensor to the plug of the pulse oximeter.
2.
Connect the plug of the pulse oximeter at the rear of the device.
To view the oximetry values, from the
Monitoring
menu, select
Monitoring
.
Adding supplemental oxygen
Oxygen may be prescribed by your clinician.
Note:
Up to 30 L/min can be added.
WARNING
•
Oxygen flow must be turned off when the device is not operating, so that unused oxygen does
not accumulate within the device and create a risk of fire.
•
Use only certified, clean oxygen sources.
•
ResMed strongly recommends adding oxygen into Stellar's oxygen inlet at the rear of the device.
Entraining oxygen elsewhere, ie into the breathing system via a side port or at the mask, has
potential to impair triggering and accuracy of therapy/monitoring and alarms (eg, High Leak
alarm, Non-vented mask alarm). If used in this way, therapy and alarm operation must be verified
each time oxygen flow is adjusted.
•
Oxygen supports combustion. Oxygen must not be used while smoking or in the presence of an
open flame. Only use oxygen supply in well-ventilated rooms.
•
The breathing system and the oxygen source must be kept at a minimum distance of 2 m away
from any sources of ignition (eg, electrical devices).
•
Oxygen must not be used while the device is being operated within the mobility bag.
To add supplemental oxygen
Refer to illustration I.
1.
Unlock the low flow oxygen inlet on the back of the device by pushing up on the locking clip.
Summary of Contents for Stellar 100
Page 212: ...2 1 I 2 1 J 1 2 3 4 5 K ...