Ventilation modes
English
73
Parameter
Setting
Resp.rate
(per min)
Mandatory breaths:
Adult: 2 to 50 [15]
Paed: 5 to 80 [15]
PEEP
(cmH
2
O)
Off, 3.0 to 20.0 [5.0]
PS
(cmH
2
O)
Spontaneous breaths:
Adult: 2 to 50 [7]
Paed: 2 to 50 [7]
Vt
(Tidal Volume) (mL)
Mandatory breaths:
Adult: 100 to 2,500 [500]
Paed:50 to 300 [100]*
PIF
(L/min)
When Volume Breath option is set to PIF
Mandatory breaths:
Adult: 10 to 120 [50]
Paed: 5 to 60 [10]
Ti
(Inspiratory time) (sec)
When Volume Breath option is set to Ti
Mandatory breaths:
Adult: 0.3 to 3.0 [1.0]
Paed: 0.3 to 3.0 [0.6]
Flow shape
(%)
Mandatory breaths:
100 (Constant), 75, 50, 25 [100]
Cycle
(%)
Spontaneous breaths:
5 to 90, Auto [Auto]
Trigger type
Flow/Pressure
Trigger
When Trigger Type is set to
Flow:
(double limb circuit only)
Adult: 0.5 to 15 [1.0] (L/min)
Paed: 0.5 to 15 [0.5] (L/min)
When Trigger Type is set to
Pressure:
(double and single limb circuit)
Very Low to Very High [Medium]
Rise Time
(msec)
Spontaneous breaths:
Min, 150 to 900 [200]
*The International ventilator standard indicates that Paediatric patient type is intended to be used for a patient receiving less
than 300 mL, however Astral permits adjustment of ‘Vt’ setting parameter up to 500 mL for cases where ‘Vt’ is set such that it
compensates for leak in the breathing circuit.
WARNING
ResMed does not recommend 500 mL as the upper limit for the pediatric tidal volume use;
however, clinicians may choose this upper limit based on their clinical determination.