Page 6
2C
Clean the cartridge bottom.
Grab Here
With 1 hand, hold the cartridge barrel and clean the cartridge bottom with an alcohol
wipe.
●
Do not
remove or rotate the cartridge top or bottom.
●
Do not
touch the bottom of the cartridge after cleaning with alcohol wipe.
2D
Load the cleaned cartridge into the on-body infusor and firmly press on the top
until it is secured in place. Make sure that you give your injection
within 5
minutes
after loading the cartridge.
Do not
insert the cartridge more than
5
minutes before injection. This can dry out the medicine.
Load
cartridge
straight
Press
down
firmly
Inject within
5 minutes
after loading
the cartridge
Insert the cartridge bottom first.
●
Do not
touch the start button until you have placed the loaded on-body infusor
on your skin.
5 minutes