Installation
Fitting VPCP and VMCP to the MRS / MRS2 rack
The VPCP and VMCP heated change ports can be fixed to an MRS or MRS2 rack system. It is recommended that they are attached to the MRS
/ MRS2 rail using the following procedure, where it is assumed that the MRS / MRS2 rack system is correctly installed.
1. Insert one of the fixing screws through the VPCP / VMCP.
2. Position the VPCP / VMCP underneath the rail and locate the respective T-nut within the rail*.
3. Hand tighten the fixing screw into the T-nut and repeat the process for the next fixing screw.
4. Position the VPCP / VMCP and tighten both fixing screws using the hexagonal key supplied.
NOTE:
The image above shows the VPCP change port but the procedure is the same for the VMCP change port.
* NOTE:
T-nuts must be used with the MRS system. However T-nuts and D-nuts are compatible with the MRS2 system.
NOTE:
Please refer to the REVO-2 change port system spacing guide (Renishaw part number H-1000-5408) for full details of the
recommended spacing for positioning all sensors and artefacts on the MRS / MRS2 rail.
RVP user's guide
www.renishaw.com
Issued 06 2022
19