RMP600 installation guide
4.4
Trig
g
er Logic
Changing the probe settings
Insert the batteries or, if they have already been installed, remove them for five seconds and then refit
them.
Following the LED check, immediately deflect the stylus and hold it deflected until five red flashes have
been observed (if the battery power is low, each red flash will be followed by a blue flash).
Keep the stylus deflected until the
“Switch-on method”
setting is displayed, then release it. The probe
is now in configuration mode and Trigger Logic™ is activated.
CAUTION:
Do not remove the batteries whilst in configuration mode. To exit, leave the stylus
untouched for more than 20 seconds
Key to the symbols
LED short flash
LED long flash
Deflect the stylus for less than 4 seconds
to move to the next menu option.
Deflect the stylus for more than
4 seconds to move to the next menu.
To exit, leave the stylus untouched for
more than 20 seconds.
LED check
Switch-on method
Radio on
(omitted if
“multiple probe mode”
was selected)
Shank on
Spin on
1
> 5 s
2
3
Battery status
Battery good
or
Battery low
Switch-off method
(omitted for shank on)
Radio off or
Spin off
Short timeout
12 s
Medium timeout
33 s
Long timeout
134 s