Renishaw RGH34 Installation Manual Download Page 10

Readhead set-up

When mounting the readhead, ensure that the scale, readhead optical window and mounting 
surfaces are clean and free from obstructions.

NOTE: 

Refer to ‘Maintenance and cleaning’ section of this manual for cleaning instructions. 

Adjusting set-up

Once the readhead is positioned, the rideheight, pitch, yaw and roll of the unit may need to be  
adjusted to achieve optimum signal strength. It should be noted that the readhead and interface  
must be properly connected to the power supply to enable the set-up LED to be used. 

Confirming set-up

For reliable operation, the LED should be Green when the readhead is moved slowly  
(<1 m/s) along the full axis travel. 

NOTE: 

The set-up LED will not indicate an optimised set-up over the reference mark.  

For further details, refer also to ‘Reference mark set-up’.

The set-up LED will exhibit one of three colours during normal use;

Green Orange  Red

Yaw 

0° ±1°

Pitch 

0° ±1°

Roll 

0° ±1°

Rideheight 

1.5 mm ±0.15 mm

Set-up tolerances

RGH34 RGS40

 

installation guide

8

Reference mark set-up

To ensure unidirectional repeatability, the reference mark requires phasing with the scale in the direction of  
normal datuming operation. A reference pulse is output in both directions, but repeatability is guaranteed only  
in the phased direction. Ensure readhead is set up correctly with a green LED indication over the full length of travel 
and that the reference mark actuator is fitted correctly.

NOTE: 

It is recommended that a datum procedure is performed as part of any power-up sequence to ensure the  

correct datum position is recorded.

NOTE: 

Reference mark output is synchronised with the incremental channels, giving unit of resolution pulse width.

NOTE: 

Check that the reference mark sensing option has been specified on the RGH34 and RGI34 that you  

have purchased.

Phasing procedure

The readhead must be moved over the reference mark in the direction to be used for the datuming operation.  
The reference mark is phased correctly when the set-up LED flashes Red for 0.25 seconds. If the set-up LED  
flashes Orange or goes Blank, the reference mark adjuster screw should be turned anti-clockwise by 

1

 

8

turn  

and the procedure repeated until a Red flash is obtained.

NOTE: 

If the readhead mounting is disturbed at any time, the reference mark will have to be re-phased.  

A reference pulse is output in both directions, but because the reference mark can only be phased for  
one direction of traverse, any indication from the LED when the readhead is  
moved in the reverse direction is to be ignored. LED flash duration  
is a constant 0.25 seconds, regardless of traverse speed.  
LED repeater (external LED driver) signals will mimic  
the on-board LED apart from the ‘Blank’ condition.  
The repeater LED will go Blank when the  
incremental output is in error (3-state).

Red  Orange   Blank

LED flash during reference mark 

traverse only

1

 

8

turn

Summary of Contents for RGH34

Page 1: ...Installation guide M 9537 9024 01 A RGH34 RGS40 linear encoder system...

Page 2: ...ion drawings 5 Scale application 6 End clamps 6 Reference mark and limit switch actuator installation 6 Readhead mounting installation 7 Readhead set up 8 Reference mark set up 8 Limit switch 9 Output...

Page 3: ...plc Disclaimer RENISHAW HAS MADE CONSIDERABLE EFFORTS TO ENSURE THE CONTENT OF THIS DOCUMENT IS CORRECT AT THE DATE OF PUBLICATION BUT MAKES NO WARRANTIES OR REPRESENTATIONS REGARDING THE CONTENT REN...

Page 4: ...OCH3 Chlorinated Solvents CH3 CH2 5CH3 CH3CHOHCH3 N heptane Propan 2 ol Minimum bend radius RGS40 100 mm NOTE Ensure self adhesive tape is on the outside of the bend radius 95 relative humidity non co...

Page 5: ...tic reference or limit switch sensor 2 1 6 5 5 15 10 2 10 5 10 5 1 5 7 7 5 1 6 2 55 Yaw tol 1 0 2 0 09 7 5 14 4 0 8 0 5 15 Forward direction of readhead relative to scale Reference mark actuator A 953...

Page 6: ...performance hole dia 1 mm FPC cable to RGH34 readhead Available in 50 100 and 150 mm lengths Component area Set up LED 3 5 mounting holes 2 off Bracket omitted from this view 3 5 alternative mounting...

Page 7: ...ength ML 40 10 2 8 5 14 8 7 2 2 Measuring length ML Reference mark actuator position range Limit switch actuator position range 0 5 0 2 100 F axis of motion F 0 5 0 2 100 F axis of motion F 6 RGS40 S...

Page 8: ...or and if necessary adhere remaining scale manually Apply firm finger pressure via a clean lint free cloth along the length of the scale after application to ensure complete adhesion 10 Clean the scal...

Page 9: ...the cylindrical body to prevent distortion of the optics An alternative arrangement is to mount the readhead in a through hole bracket whilst using a 1 5 mm shim to maintain ride height tolerance Agai...

Page 10: ...ation over the full length of travel and that the reference mark actuator is fitted correctly NOTE It is recommended that a datum procedure is performed as part of any power up sequence to ensure the...

Page 11: ...vailable with a variety of different clocked outputs Customers must ensure they comply with the lowest recommended counter input frequency Speed Digital interfaces Non clocked output interfaces Analog...

Page 12: ...dard RS422A line receiver circuitry Capacitors recommended for improved noise immunity Customer electronics 120 A B Z Q Cable Z0 120 Interface A B Z Q 0 V 0 V 220 pF 220 pF Remote LED driver outputs T...

Page 13: ...Vpp V0 V0 Synchronised pulse Z duration as resolution S Repeatability of position uni directional maintained within 20 C from temperature at time of phasing and for speeds 0 5 m s Actuation device A...

Page 14: ...x 103 cycles at 5 mm bend radius Scale specifications Scale type RGS40 S Reflective gold plated steel tape with protective lacquer coating Adhesive backing tape allows direct mounting to the machine s...

Page 15: ...ol used in the RENISHAW logo are registered trade marks of Renishaw plc in the United Kingdom and other countries apply innovation and names and designations of other Renishaw products and technologie...

Reviews: