TP7M EP (enhanced performance)
Sense directions
6‐way: ±X ±Y Z
3D accuracy (test to ISO 10360 Pt 2)*
<0.6 μm ﴾0.000024 in﴿
Unidirectional repeatability ﴾2σ μm﴿
Trigger level 1:
0.25 μm ﴾0.00001 in﴿
Trigger level 2:
0.25 μm ﴾0.00001 in﴿
XY (2D) form measurement deviation
Trigger level 1:
±0.25 μm ﴾0.00001 in﴿
Trigger level 2:
±0.5 μm ﴾0.00002 in﴿
XYZ (3D) form measurement deviation
Trigger level 1:
±0.5 μm ﴾0.00002 in﴿
Trigger level 2:
±1 μm ﴾0.00004 in﴿
Trigger force (at stylus tip)
XY plane:
0.02 N
Z axis:
0.15 N
Overtravel force
XY plane:
0.78 N
Z axis:
11.75 N
Weight
85 g (3 oz)
Maximum extension (on a PH10 series head)
200 mm (7.87 in)
Maximum recommended stylus length (M4 styli range)
50 mm (5.9 in) steel
80 mm (7.08 in) GF
Mounting method
Multi-wired autojoint
Suitable interface
PI 7-3
Automatic probe changing
Autochange rack
Operating temperature range
10 °C ‐ 40 °C ﴾50 °F ‐ 104 °F﴿
The data above applies for test conditions as follows:
Stylus length 50 mm (1.97 in)
Stylus velocity 240 mm / min (1.57 ft / min)
* Test performed on a CMM specification U3 = 0.48 μm + L / 1000
TP7 user's guide
www.renishaw.com
Issued 10 2019
9