Renfert Silent TS2 Instruction Manual Download Page 4

- 3 -

2.2  Improper Use

Fire promoting, easily flammable or explosive materials must not be extracted with the Silent TS2.

The extraction of liquids, smoldering or burning materials is prohibited. 

This device is not intended for private, household use. 

Any use other than specified in these instructions is deemed improper and constitutes a misuse of the 

device. The manufacturer shall not be liable for damages caused by improper use. 

Only spare parts and accessories supplied or authorized by Renfert GmbH may be used with this 

product. If other spare parts or accessories are used, this could have a detrimental effect on the safety 

of the device, increase the risk of serious injury and lead to damage to the environment or the device 

itself.

2.3  Ambient Conditions for Safe Operation

The device may only be operated:

•  Indoors

•  Up to an altitude of 2,000 m above sea level,

•  At an ambient temperature of between 5  –  40 ºC [41  –  104 ºF] *),

•  At a maximum relative humidity of 80 % at 31 ºC [87.8 ºF], dropping to a linear of up to 50 % 

•  relative humidity at 40 ºC [104 ºF] *),

• 

With mains power where the voltage fluctuations do not exceed 10 % of the nominal value,

•  Under contamination level 2 conditions,

•  Under over-voltage category II conditions.

*)  Between 5 - 30 ºC [41 - 86 ºF] the device can be operated at a relative humidity of up to 80 %. At temperatures between 

31 - 40 ºC [87.8 - 104 ºF] the humidity must decrease proportionally in order to ensure operational readiness (e.g. at 

35 ºC [95 ºF] = 65 % humidity, at 40 ºC [104 ºF] = 50 % humidity). The device may not be operated at temperatures above 

40 ºC [104 ºF].

2.4  Ambient Conditions for Storage and Transport

For storage and transport the following specifications to ambient conditions apply:

•  Ambient temperature - 20  –  + 60 ºC [- 4  –  + 140 ºF].

•  Maximum relative humidity 80 %.

2.5  Hazard and Warning Information

   

2.5.1  General Information

 

If the device is not used in compliance with the supplied instructions, the safety of the device 

can no longer be guaranteed. 

 

► The device may only be operated using a mains cable with the country-specific plug system. 

Any necessary alterations must be carried out by a qualified electrician.

 

► The device may only be operated if the information on the identification plate conforms to the 

specifications of your local mains power supply. 

 

The device may only be plugged into outlets which are connected to the protective conductor 

system.

 

The mains plug must be easily accessible.

 

Disconnect the device from the mains before carrying out work on the electrical parts. 

 

Check connection cables (such as power supply cords), tubes and housing (i.e. the key-pad) 

regularly for damage (i.e. kinks, cracks and porosity) or signs of ageing.  

Devices with damaged connection cables, tubes or housing parts or other defects must not be 

operated!

 

Defective devices must be put out of service immediately. Remove the mains plug and ensure 

the device is not used. Send the device for repair!

 

Only operate the device under supervision. 

 

Please observe the national accident prevention regulations!

 

It is the responsibility of the operator that national regulations during operation and regarding a 

repeated safety inspection of electrical equipment are complied with. 

For Germany these are BGV A3 in relation with VDE 0701-0702.

EN

Summary of Contents for Silent TS2

Page 1: ...uction manual Mode d emploi Istruzioni d uso Instrucciones de servicio Инструкция по эксплуатации Instrukcja obsługi 操作说明书 사용 설명서 Silent TS2 Nr 2930 0050 2930 1050 21 6670 12112015 A Made in Germany Ideas for dental technology ...

Page 2: ...ing 7 4 2 Setup 7 4 2 1 External Exhaust Air Route 7 4 3 Electrical Connection 7 4 4 Compressed Air Connection 7 4 5 Installing the Pinch Valves 8 4 5 1 Suction Pipe Alteration 8 4 6 Connection to the Extraction Point 8 4 7 Electrical Connection 9 5 Operation 9 5 1 Switching the Unit On 9 5 1 1 Stand By 9 5 2 Operation 10 5 2 1 Extraction 10 5 2 2 Extraction force 10 5 3 Programming 11 5 3 1 Self ...

Page 3: ...equence You will be asked to carry out the action in a specified order Result of an action reaction of the device reaction of the program The unit or program reacts as a result of your actions or when a specific incident occurs Other symbols are explained as they occur 2 Safety 2 1 Intended Use This device is designed to extract dry non explosive dust This device is intended solely for commercial ...

Page 4: ...t be operated at temperatures above 40 ºC 104 ºF 2 4 Ambient Conditions for Storage and Transport For storage and transport the following specifications to ambient conditions apply Ambient temperature 20 60 ºC 4 140 ºF Maximum relative humidity 80 2 5 Hazard and Warning Information 2 5 1 General Information If the device is not used in compliance with the supplied instructions the safety of the de...

Page 5: ...ps must be taken to ensure that the exhaust air is properly venti lated Please refer to the associated safety data sheets for specific requirements Dispose of extracted material according to local statutory regulations 2 6 Authorized Persons Operation and maintenance of the device may only be performed by qualified personnel Minors and pregnant women may only operate and service the device if they...

Page 6: ... sandblaster etc Recommended use of the suction hoses WB WB Suction hoses A and B simultaneously WB WP Suction hoses A and B simultaneously WP WP Suction hose A or B alternatively D D Suction hose A or B alternatively WB D Suction hose A or B alternatively WP D Suction hose A or B alternatively 3 2 Components and Functional Elements 1 Silent TS2 2 Key pad 3 On off switch 4 Front panel Fine filter ...

Page 7: ...ady for us in the dust drawer 1 Compressed air tube sealed 2 Pinch valves with suction port 1 Connection set 2 Shockproof plug only for 2930 0050 1 4 m Suction Hose 4 PVC end bushings 3 4 Accessories 2921 0002 Dust bags 5 pieces 2921 0001 Set of rollers 2925 0000 Extractor clamp without glass screen 2925 1000 Glass screen with holder for the extractor clamp 2921 0003 Set of mufflers 2 pieces 2921 ...

Page 8: ...create negative pressure within the room which when using an air dependent naked flame fed by gas liquid or solid fuel can cause poisonous gasses e g carbon monoxide to be drawn into the working area It is therefore essential to ensure that the fresh air supply is sufficient and that the environmen tal air pressure is maintained this should then be monitored by an authorized specialist e g a certi...

Page 9: ...ng and then at the back remove from the extractor connection fitting A Lever the suction pipe out of the holder B turn C and replace back into a new position in the holder D Insert the suction pipe into the extractor connection fitting on the extraction device hang E the holder into the holding screws 26 Fix the holder with the knurled head screw 19 Push the pinch valve into the suction pipe as fa...

Page 10: ... with protective contact is available for use as an adapter This adapter may only be put together by an electrician The protective conductor system must not be interrupted by the adapter 5 Operation The extractor unit is operated via the buttons on the key pad 2 Fig 1 5 1 Switching the Unit On The extractor is switched ON and OFF at the On Off switch 3 When the unit is switched on a self diagnosis...

Page 11: ...educes extraction force Readjustment of the extraction force The set extraction force applies for both suction hoses When another suction hose is opened or closed the power in the suction turbine is automatically readjusted so that the selected extraction force may be achieved in each opened suction hose To deactivate the extraction force readjustment In the stand by mode Press buttons 43 41 toget...

Page 12: ...in chapter 6 5 5 3 1 Self Diagnosis The diagnosis symbol 31 shows whether the self diagnosis has been activated or not Press menu key 30 1 x The diagnosis symbol 31 blinks Switch the self diagnosis ON or OFF using the buttons 41 43 Press the enter button 40 5 3 2 Audible Signal Various settings and warnings are confirmed or alerted via an audible signal This signal can be switched on or off The au...

Page 13: ...hold the point at which the connected electrical appliance opens the suction hose can be individually adjusted for each hose Press menu key 30 4 x The start up threshold symbol 34 blinks The letters A and B blink Press the suction hose button 39 of the suction which is to be adjusted The letter of the selected hose is indicated The display shows OFF Switch the electrical appliance off Units equipp...

Page 14: ...me in order to prevent short interruptions e g when working with a hand piece motor from switching the extractor on and off too frequently therefore causing the suction power to be continuously be readjusted Press the menu key 30 5 x The shut off delay symbol 35 blinks Set shut off delay A The letter A is indicated The display shows the current shut off delay setting for suction hose A Set the req...

Page 15: ...t bag into the dust drawer Ensure that the new dust bag is properly aligned in the guide grooves 21 and the glue flap faces upwards Use only original Renfert dust bags see spare parts list Inspect the dust drawer seal for damage and replace it as required 24 Insert the dust drawer making sure it is straight and push it in as far as it will go Make sure that the dust drawer is seated correctly on t...

Page 16: ...te so that the smooth compressed side of the filter mat faces down or outwards when the filter is installed Replace the new filter cassette with the new filter Make sure the filter cassette is properly seated and locked in place on all lock tabs 6 2 3 Replacing the Electronics Filter Cooling air for the unit s electronics passes through the electron ics filter 11 The type of material trapped in th...

Page 17: ...ritten by factory set values Extraction off and Press button 41 43 and keep pressed Extraction on and Press button 41 43 and keep pressed An audible signal will confirm the factory set parameter is saved and the display will indicate rES Factory set parameter Function Feature Adjustment Range Factory Parameter Extraction force one suction hose is open 1 3 2 Extraction force both suction hoses are ...

Page 18: ...e is too high Switch the extractor on Chapter 5 1 Check and reinstall the fuse if necessary Chapter 6 4 Connect the dust generating device to the coupler socket Chapter 4 7 Adjust start up threshold Chapter 5 3 4 Extractor starts al though a connected de vice is not being used Start up threshold of one suction hoses is too low Check which suction hose is open the relevant letter will be indicated ...

Page 19: ...epair service Error Code Cause Meaning Corrective Action Cod01 The set dust bag fill level has been reached Change dust bag Chapter 6 1 Err01 Suction flow is blocked Insufficient compressed air pressure Check the suction hoses for blockages examine the dust bag fill level and change if necessary See chapter 6 1 examine the fine particle filter for contamination and change if necessary see chapter ...

Page 20: ...t max volume flow 59 dB A Number of suction hoses 2 Volume flow max per suction hose 3300 l min Vacuum pressure max 20 kPa 2 9 psi Fine particle filter Filter surface area ca 0 8 m ca 1240 sq inch FIlterquality Class M according to EN60335 2 69 Fill volume dust bag ca 7 5 l 2 US Gallonen Dimensions width x height x depth 225 x 595 x 565 mm 8 8 x 23 4 x 22 2 inch Weight empty ca 27 kg ca 59 5 lbs Ø...

Page 21: ...ll not extend the original warranty 10 Disposal Information 10 1 Disposing of Consumables Full dust bags and filters must be disposed of under compliance with locally applicable regulations Depending on the material trapped by the filters protective gear may need to be worn during disposal 10 2 Disposing of the Unit The unit must be disposed of by an authorized recycling operation The selected fir...

Page 22: ...nfert USA 3718 Illinois Avenue St Charles IL 60174 USA Tel 1 6307 62 18 03 Fax 1 6307 62 97 87 www renfert com info renfertusa com USA Free call 800 336 7422 Hochaktuell und ausführlich auf Up to date and in detail at Actualisé et détaillé sous Aggiornato e dettagliato su La máxima actualidad y detalle en Актуально и подробно на www renfert com Ideas for dental technology ...

Reviews: