24 -
User manual
6
L.i.o.L.
Linear Input over Limits (Word modbus 2006)
If linear input, allows process to go over limits (Par. 3 and 4).
0
Disabled (
Default
)
1 Enabled
7
o.cAL.
Offset Calibration (Word modbus 2007)
Value added / subtracted to the process visualization (usually correcting the
value of environmental temperature).
-10000..+10000 [digit
1
] (degrees.tenths for temperature sensors).
Default
0.
8
G.caL.
Gain Calibration (Word modbus 2008)
Percentage value that is multiplied for the process value (allows to calibrated
the working point). Ex: to correct the range from 0..1000°C showing 0..1010°C,
set the par. to -1.0.
-1000 (100.0%)...+1000 (+100.0%),
Default
: 0.0.
9
Ltch
Latch-On (Word modbus 2009)
Automatic setting of limits for linear inputs and potentiometers.
0
Disabled (
Default
)
1 Standard
2
Virtual zero
3
Linear virtual zero
10
FLtr.
Filter (Word modbus 2010)
Analogue input reading filter: increases process stability on word 1000, 1100,
1200, 1232.
1...50. (
Default
: 1)
11÷15
Reserved Parameters - Group A
Reserved parameters - Group A