background image

ISO-9001 Certified

        13940014-101 REV XA

SPECIFICATIONS:

Catalog

 13940014 (deg F) / 13940015 (deg C)

Reading Scales -  

EG  = Freeze point, ethylene glycol antifreeze 

  

PG  =

 

Freeze point, propylene glycol antifreeze

  SG  = Specific gravity, battery acid

  WW =  Freeze point, windshield washer solution

Reading Ranges,
  EG, 

32° thru -70°F / 0° thru -57°C,  Accuracy  +/- 0.5°F / 0.3°C

  PG, 

32° thru -60°F / 0° thru -51°C,  Accuracy  +/- 0.5°F / 0.3°C

  Battery, 

1.000 - 1.463 SG,   Accuracy  +/- 0.003 SG

  Windshield Washer,

 32° thru -65°F / 0° thru -54°C, 

        Accuracy  +/- 2°F / 1°C

Calibration -

 Distilled Water

Automatic Temperature Compensation -

 68°F (20

o

C)

The built in temperature compensation mode is based on  

the temperature coefficients that have been calculated for 

the individual solutions offered in this instrument.

Prism -

 Glass

Illumination -

 589nm LED

Dimensions -

 54 x 27 x 100 mm / 2.13 x 1.06 x 3.9 inches

Power - 

2 AAA Batteries, included.

Battery life - 

10,000 readings, Auto Off Sleep Mode.

Ratings - 

IP65 Dustproof/Water Resistant, CE, RoHS, and WEEE 

  compliant.

Warranty - 

One year against manufacturing defects. Evidence of 

     tampering voids warranty.

AUTOMATIC TEMPERATURE COMPENSATION:

All Reichert refractometers are automatically temperature 
compensated assuring extremely accurate results. Automatic 
temperature compensation corrects readings over a range of 
temperatures. As an example; samples taken within the working 
temperature range of the instrument are corrected to a reference 
temperature of 68° F (20° C). Temperature compensation is essential 
because refractive index varies inversely with temperature. This 
feature is used in lieu of temperature control of the measuring 
surface and sample. For this feature to be effective however, it is 
necessary to allow the sample to temperature equilibrate to the 
ambient temperature of the measuring surface. Without allowing for 
temperature equilibration of the sample the read values will not be as 
accurate as they could be since the sample will be in a state of 
temperature and refractive index change.

ERROR CODES:

Err01 - No sample present. Add sample.
Err02 - Inadequate sample. Add additional sample.
Err03 - Sample exceeds the refractive index reading range.
Err04 - Sample temperature has not equilibrated. Allow more time for  
            temperature equilibration.
Err05 - Excessive ambient light. Cover sample well when reading.
Err06 - Excessive ambient light. Cover sample well when reading.
Err07 - Positive calibration error. Re-calibrate with distilled water.
Err08 - Negative calibration error. Re-calibrate with distilled water.
Err09 - Poor sample condition. Sample may not be able to be read.
Err10 - Sample type may not be able to be read.
Err12 - Index of sample is out of range of instrument.
Err5X - For any errors in this series contact Reichert Analytical 
            Instruments for technical assistance.

BATTERY CONDITION:

The following display examples represent either a low battery condi-
tion or an extremely low battery condition where the instrument will 
no longer operate properly. In either case the batteries should be 
promptly replaced.

      Battery condition low. 
      Replace Batteries.

      Battery condition extremely low. 
      Replace Batteries.

Reviews: