immediately. If the device is not used in accordance with this manual or if it is used when the 
device exhibits functional differences from those stated in this manual, then our company is not 
responsible for any damage to or caused by the device. 

ƒ

 

Do not try to open, remove protective covers, or dismantle the device for any reason. There is a 
danger of electrical shock and/or serious injury. Even the replacement of the lithium battery and 
the air micro-filter must be done an authorized member of the service department only. 

ƒ

 

All material and parts which come into direct contact with the client's body (such as cleaning 
agents, the applicators, electrodes) must respectively comply with the standards related to 
irritants, allergens, toxins, geotaxis and carcinogens (ISO 10993-1, ISO 10993-3, ISO 10993 5). 
The user is responsible for all these materials and parts if they were not supplied by an 
authorized equipment supplier. 

ƒ

 

The connectors for accessories, as well as the other connectors, must not be used for 
connecting anything else other than what they are designed for. There is a danger of electrical 
shock and/or serious damage to the device. 

ƒ

 

The device does not use or emit any toxic substances during its operation, storage or transport 
under the stated conditions. 

ƒ

 

Before the start of therapy make sure that all set parameters comply with your requirements. 

ƒ

 

To terminate operation, do not use the main power switch! Instead, press the "start/stop". 

ƒ

 

The time interval between turning off the main power switch and turning it back on must be at 
least 3 seconds. 

ƒ

 

Place the device so that it is possible to disconnect the connector of the power supply adapter 
quickly and easily from the device, or the male plug of the power supply adapter from the 
mains. To disconnect from the mains, unplug the male plug of the power supply adapter from 
the mains socket outlet 

ƒ

 

If it is necessary to discard the device, the lithium battery must be removed. The removed 
battery must be disposed according to local hazardous waste disposal requirements. Do not 
place the device in municipal waste containers. The device itself does not contain any toxic 
materials which could harm the environment when disposed of ecologically 

ƒ

 

The device and its accessories must be used in compliance with this manual. 

ƒ

 

The device must be placed out of the reach of children. 

ƒ

 

The device does not contain any components, except for the fuse and parts containing in the 
shockwave transmitter replacement kit, which can be repaired or replaced by the user. Do not 
remove the cover from the control unit. An authorized service department member must do all 
repairs. 

ƒ

 

Do not disconnect an applicator during therapy.

 

 

 

 

 

Summary of Contents for APEX MAGNUM

Page 1: ...2019 APEX MAGNUM DUAL FUNCTIONING ACOUSTIC WAVE THERAPY SYSTEM Revision 1 12 11 2018 User s Guide...

Page 2: ...ion 9 2 3 Pneumatic Accessories 10 2 4 Electromagnetic Accessories 10 2 5 Replacement Parts 11 2 6 Pneumatic Transmitter Adapter Replacement 12 2 6 1 Pneumatic Piston Replacement 12 2 7 Electromagneti...

Page 3: ...ON WITH THIS DEVICE Only use BABY OIL IF DESIRED The use of gels or lotions render both hand pieces inoperable The damage from the use of gels or lotions is not covered under the warranty Contact Refi...

Page 4: ...panel USER S INSPECTION DAMAGE CAUSED IN TRANSIT The unit and all accessories must be checked for potential defects transport damage or loss immediately after receipt Specifications are subject to cha...

Page 5: ...g feature Based on detailed research and practical use of the device the well organized predefined programs will provide recommendations for the treatment of various conditions The device consists of...

Page 6: ...ive pressure amplitude 1 to 10 MPa Short time duration 1 mu to 20 mu Rapid pressure increase 100 ns Broad frequency spectrum 1 Hz to 1 MHz For therapeutic applications these values are lower especiall...

Page 7: ...corticoid therapy Shockwave application can cause undesirable heart activity 1 4 INTENDED USE FOR SHOCKWAVE TREATMENT ESWT is for the intended use as a treatment for musculoskeletal pain 1 5 CONTRAIN...

Page 8: ...ug 8 Filter cup with pressure gauge 9 Plug for pedal switch 10 Plug for Electromagnetic wave handle 2 1 ACCESSORIES LIST Device 1pc A shock wave handle 2pcs E Shock wave handle 2pcs A handle transmitt...

Page 9: ...itter Adapters Pneumatic transmitter adapter treatment heads 11 Electromagnetic Transmitter Adapters Electromagnetic transmitter adapter treatment heads 5 Foot Pedal Foot pedal for power 1 Apex Silico...

Page 10: ...to the bottom cooling vents Do not put any heat producing devices or any objects containing water or other liquids on the device Do not place the device close to appliances producing strong electroma...

Page 11: ...Transmitter Adapters B 15 mm B 20 mm B 25 mm Convexed Pneumatic Transmitter Adapters C 15 mm C 20 mm C 25 mm Concaved Pneumatic Transmitter Adapters 2 4 Electromagnetic Accessories Electromagnetic Sho...

Page 12: ...escription Pneumatic Handle Pneumatic hand piece Electromagnetic Handle Shockwave hand piece Electromagnetic Transmitter Adapters Shockwave handle tips Foot Pedal Foot pedal for power Pneumatic Transm...

Page 13: ...to choose the proper transmitter Secure the transmitter on the applicator prior to treatment 1 Transmitter Adapter Screw Cap After Installation 2 6 1 Pneumatic Transmitter Piston Replacement Assembled...

Page 14: ...ols to choose the proper transmitter Secure the transmitter on the applicator prior to treatment Unscrew adapter by twisting counter clockwise until removed from handpiece Screw chosen adapter head in...

Page 15: ...ng to select language set time Notice ADVANCED SETTING is only for Engineering debugging DO NOT USE Click A shock to enter Interface as FIG 3 1 Click on the body part chosen for treatment on the scree...

Page 16: ...f rounds completed 600 impacts are one round 10 Frequency 6 HZ stands for the system default work frequency is 6 HZ Energy 2 0 BAR stands for the system default work Energy is 2 0 BAR click on and to...

Page 17: ...the default intensity frequency and number of impacts The automatic mode can also be adjusted freely to the most suitable parameters Treatment example for Elbow adipose tissue In interface as Fig 4 1...

Page 18: ...15HZ the energy is 50 130MJ When the frequency is 16HZ the energy is 50 100MJ 3 3 Touch Screen Calibration 1 Before powering on the device press on the screen with your finger do not remove your finge...

Page 19: ...nd piece control board o Please check if the plug of the cooling handle 8 is connected accurately Reboot device E1 Connection failure between display control board and function control board This may...

Page 20: ...2 detergent solution to clean the exterior of the Apex Therapy System and its parts Never use cleaning agents containing alcohol ammonia benzene thinners etc Never use abrasive cleaning materials whi...

Page 21: ...lled and tested according to the existing valid standards IEC 364 If it is not known whether cables are safe and or correct then they should be checked and or upgraded by an inspection engineer Check...

Page 22: ...ransport under the stated conditions Before the start of therapy make sure that all set parameters comply with your requirements To terminate operation do not use the main power switch Instead press t...

Page 23: ...ents are not covered Failure to do so may result in permanent damage Correct usage of protective sheath Correct usage of protective sheath Improper use of protective sheath Sheath covering air vents C...

Page 24: ...gether with a photocopy of your bill of sale What Refine USA LLC will do Refine USA LLC will evaluate your report of a possible defect to determine whether a defect exists and if it does exist Refine...

Reviews: