Instruction Manual TAP, TAP-CLAMP / Temperature And Pressure Gauge English
25
2. Turn off the clamp.
3. While holding [UNIT], press [ON / OFF] until display shows "CALL"
4. Release both buttons, the display shows "00"
5. Press [UNIT] button < 1 sec. to adjust the number to "06"
6. Press [ON / OFF] button < 1 sec., the display shows "00 °C"
7. Press [UNIT] button < 1 sec. to show a code of temperature
8. Put the K – thermocouple (plate) into 0 °C ice water.
9. Wait the temperature code reading to be steady, press [UNIT] button < 1 sec.
10. The display shows the ambient temperature.
11. Press [UNIT] button < 1 sec. to quit the temperature calibration.
Room temperature calibration
1. Turn off the clamp.
2. Place the clamp at constant temperature at 25
3 °C condition for 2 hrs. to ensure
the temperature of the K – thermocouple to be even.
3. While holding [UNIT], press [ON / OFF] until display shows „CALL“
4. Release both buttons, the display shows „00“
5. Press [UNIT] button < 1 sec. to adjust the number to „09“
6. Press [ON / OFF] button < 1 sec. to show a code of temperature.
7. Wait the temperature code reading to be steady, press [ON / OFF] button < 1 sec.
8. The display shows the ambient temperature.
9. Press [ON / OFF] button < 1 sec. to quit the temperature calibration.