Calibration

1.  Open the cover plate and place 2-3 drops of distilled water onto the prism 

surface, close the cover plate, press lightly and ensure the water spreads 
evenly across the whole surface area without air bubbles or dry sections. 
(wait 30 seconds until the water adjusts to ambient temperature)

2.  Hold towards light source and look into the eyepiece, if the boundary line 

between the upper blue field and the lower white field is not located exactly 
on the 0°C waterline, adjust the correcting screw until both lines meet.

3.  Wipe the water from the prism with clean dry cloth and replace the rubber 

correcting screw cover.  

NOTE

: under normal operating conditions the calibration only needs to be checked 

periodically.

Maintenance

1. 

If the ambient temperature is not within the automatic temperature  
compensation range (10 - 30°C) the instrument must be calibrated before 
testing commences.

2. 

The prism must be cleaned thoroughly with a soft, clean damp cloth after 
each use - failure to do so will result in inaccurate readings and possible 
long term  
damage.

3. 

The instrument must be kept in a dry clean environment to prevent  
condensation forming on the optical components.

4. 

Use a glass rod to sample battery fluid as acid will damage the dropper and 
also harm your skin–separate readings must be taken from each cell

5. 

If used in accordance with the instructions and all warning are followed - 
the optical performance will remain constant and the instrument will give 
many years of reliable service.