Specifications

Measuring Range: 

-50 to 0°C propylene and ethylene glycol  

 

1.15 - 1.30 specific gravity 

Minimum Scale Division: 

5°C

Dimensions: 

30 x 40 x 170mm (1 x 1.5 x 6.5”)

Automatic Temperature  Compensation Range: 
 

10 to 30°C            

Weight: 

200g (7 oz)

Instrument Description 

1. Prism
2.  Cover Plate
3.  Correcting Screw
4.  Mirror Tube
5.  Eyepiece (Adjusting ring of diopter)

Operating Instructions

1.  Use dropper to obtain coolant sample, open the cover plate and place 2-3 

drops of the sample onto the prism surface, close the cover plate, press 
lightly and ensure the water spreads evenly across the whole surface area 
without air bubbles or dry sections. Wait 30 seconds until the liquid sample 
adjusts to ambient temperature.

2.  Hold towards light source and look into the eyepiece to view the scale, the 

coolant readings are located on the far left and right and the battery fluid 
reading is in the centre.

3.  The test reading is taken from where the boundary line between the upper 

blue field and the lower white field crosses the graduated scale. This reading 
is the actual freezing point/concentration level of the coolant or the charge 
state of the battery fluid.

4.  Wipe the prism clean and dry with damp soft cloth and store in the supplied 

case.

 