7

Auto Power OFF

To preserve battery life, the meter is programmed to turn itself off after 30 
minutes of inactivity. To turn this function off, while the meter is off, press 
and hold the Time Weighting Button while turning the meter on. When the 
meter turns on release both the Time Weighting Button and Power button. 
The Auto Power Off indicator will not appear on the display, confirming 
this feature is now disabled. If the meter is turned off then back on, the 
"Auto Power Off" feature will be enabled again.

Calibration Procedure

In order to calibrate the R8060 an external calibrator that can provide a 

94.0dB signal is required (REED R8090). If the meter has been exposed 
to extreme conditions or has not been used in a long period of time the 
meter should be calibrated by following steps 1 through 6 below.

1.  Remove windshield ball.
2.  While the meter is off, press and hold the MIN /MAX Hold button 

while turning the meter on.

3. 

When the meter turns on, release both the MIN/MAX Hold and Power 
button to enter calibration mode as indicated on-screen by "CAL 94dB".

4.  Insert the microphone into the calibrator.
5.  Press the UP and DOWN arrow buttons to increase or decrease the 

dB level to match the Sound Calibrator output.

6.  Once the desired dB level has been attained press the 

MIN/MAX

 

button to save the setting and exit the calibration mode.

NOTE

: Press the Power button to exit the calibration mode at any time.

OFF

94dB

114dB

(1KHz)

IEC 942 CLASS 

2

R8090

Sound Level Calibrator

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for R8060

Page 1: ...Sound Level Meter Instruction Manual R8060 www com information itm com 1 800 561 8187...

Page 2: ...C DC Signal Output Jack 8 Applications 9 Accessories and Replacement Parts 9 Product Care 9 Product Warranty 10 Product Disposal and Recycling 10 Product Support 10 Introduction Thank you for purchasi...

Page 3: ...ther than for the purpose of replacing batteries may cause damage that will not be covered under the manufacturer s warranty Servicing should only be provided by an authorized service center Features...

Page 4: ...enser Digital Analog Bargraph Yes 30 segment Display 4 digit LCD Min Yes Max Yes Alarm Indicators Under and Over visual on screen Auto Shut off Yes after 30 mins off Tripod Mountable Yes Low Battery I...

Page 5: ...0 AC DC Output Terminal 11 Tripod Mounting Screw 12 Battery Cover Display Description Bar Graph Indicator Reading Indicator Battery Capacity Indicator C Weighting Indicator Auto Power Off Indicator Un...

Page 6: ...upper limit of the selected dB range has been exceeded Adjust the selected dB range until the over range indicator disappears from the screen Further more if the under range indicator appears during y...

Page 7: ...been exposed to extreme conditions or has not been used in a long period of time the meter should be calibrated by following steps 1 through 6 below 1 Remove windshield ball 2 While the meter is off p...

Page 8: ...external systems For example the AC output can be used with a frequency analyzer or oscilloscope to identify the frequency level during a measurement The DC output is used to identify the dB level du...

Page 9: ...your part listed here For a complete list of all accessories and replacement parts visit your product page on www reedinstruments com Product Care To keep your instrument in good working order we rec...

Page 10: ...at 1 877 849 2127 or by email at info reedinstruments com to discuss the claim and determine the appropriate steps to process the warranty Product Disposal and Recycling Please follow local laws and...

Page 11: ...Test Measure with Confidence Check out our latest Products www com information itm com 1 800 561 8187...

Page 12: ...AIR VELOCITY TEMPERATURE HUMIDITY SOUND MOISTURE ELECTRICAL www com information itm com 1 800 561 8187...

Reviews: