REED R6001 Instruction Manual Download Page 7

7

www.REEDINSTRUMENTS.com

3.  When the LCD flashes, press the 

REL

 button to enter calibration 

mode. This will be indicated by "CAL 1" on the LCD.

4.  Insert the tip of the meter into the 33%RH Humidity Calibration Standard 

(REED R9933) then press the 

MAX/MIN

 button to start the adjustment. 

The calibration adjustment will be confirmed by a long beep.

5.  The meter will then display "CAL 2" to calibrate the second calibration 

point. Insert the tip of the meter into the 75%RH Humidity Calibration 

Standard (REED R9975) then press 

MAX/MIN

 button to start the adjust-

ment. The calibration adjustment will be confirmed by a long beep.

6.  After both calibrations are complete, the meter will resume normal 

operation.

Note:

 The calibration steps can be cancelled at any point by pressing the 

ON/OFF button.

Restoring Factory Settings

1.  Turn the meter off.
2.  Press the HOLD, °C/°F, and power button simultaneously, then release.
3.  When the LCD flashes, press the 

MAX/MIN

 button to restore factory 

default settings.

4.  A long beep will confirm factory settings have been restored. The 

meter will then resume normal operation.

Battery Replacement

When the low battery indicator appears on the display, the battery should be 
replaced. Remove the battery cover on the back and insert a new 9V battery.

Accessories and Replacement Parts

CA-05A

 Soft Carrying Case

R9933

 33% Humidity Calibration Standard

R9975

 75% Humidity Calibration Standard

R9980

 33% and 75% Humidity Calibration Kit

Don't see your part listed here? For a complete list of all accessories and 
replacement parts visit your product page on www.reedinstruments.com.

1.800.561.8187 

[email protected]

REED-Direct.ca

Summary of Contents for R6001

Page 1: ...www REEDINSTRUMENTS com Thermo Hygrometer Instruction Manual R6001 1 800 561 8187 info REED Direct ca REED Direct ca ...

Page 2: ...umidity Calibration Procedure 6 7 Restoring Factory Settings 7 Battery Replacement 7 Accessories and Replacement Parts 7 Applications 8 Product Care 8 Product Warranty 8 Product Disposal and Recycling 9 Product Support 9 Introduction Thank you for purchasing your R6001 Thermo Hygrometer Please read the following instructions carefully before using your instrument By following the steps outlined in...

Page 3: ...or the purpose of replacing batteries may cause damage that will not be covered under the manufacturer s warranty Servicing should only be provided by an authorized service center Features Easy to use intuitive design Dual LCD readout displays temperature and humidity User selectable F C High Sampling Rate of 2 5 times sec Data hold Min Max and Relative functions Low battery indicator and auto shu...

Page 4: ...4 to 140 F 20 to 60 C Storage Temperature 14 to 140 F 10 to 60 C Operating Humidity Range 10 to 80 Dimensions 9 5 x 2 1 x 1 3 240 x 54 x 34mm Weight 6 4oz 180g Included Carrying Case Battery Instrument Description 1 Sensor Protective Cap 2 LCD Display 3 Power Button 4 MAX MIN Button 5 Data Hold Button 6 Relative Function Button 7 Unit of Measure Button 8 Battery Cover 1 2 3 4 5 6 7 8 1 800 561 818...

Page 5: ...FF Unit of Measure Selection Temperature When the meter is powered on press the C F button to toggle between Celsius and Fahrenheit Note The default unit of measure is Celsius C The meter will automati cally save the last selected unit of measure when turned off Data Hold Press the HOLD button to freeze the current reading on the display Press the button again to resume normal operation Note When ...

Page 6: ... through stored measurements 5 To exit MAX MIN mode and resume normal operation press and hold the MAX MIN button for two seconds Note When in MAX MIN mode the REL and C F buttons are disabled Auto Power Off To preserve battery life the meter is programmed to turn itself off after 30 minutes of inactivity To disable this feature press and hold the HOLD button while powering on the meter When the m...

Page 7: ...steps can be cancelled at any point by pressing the ON OFF button Restoring Factory Settings 1 Turn the meter off 2 Press the HOLD C F and power button simultaneously then release 3 When the LCD flashes press the MAX MIN button to restore factory default settings 4 A long beep will confirm factory settings have been restored The meter will then resume normal operation Battery Replacement When the ...

Page 8: ...free of defects in material or workmanship for a period of one 1 year from date of shipment During the warranty period REED Instruments will repair or replace at no charge products or parts of a product that proves to be defective because of improper material or workmanship under normal use and maintenance REED Instruments total liability is limited to repair or replacement of the product REED Ins...

Page 9: ...e contact your authorized REED distributor or REED Instruments Customer Service by phone at 1 877 849 2127 or by email at info reedinstruments com Please visit www REEDINSTRUMENTS com for the most up to date manuals datasheets product guides and software Product specifications subject to change without notice All rights reserved Any unauthorized copying or reproduction of this manual is strictly p...

Page 10: ...www REEDINSTRUMENTS com TEST MEASURE WITH CONFIDENCE CHECK OUT OUR LATEST PRODUCTS 1 800 561 8187 info REED Direct ca REED Direct ca ...

Reviews: