REED R6001 Instruction Manual Download Page 6

6

www.REEDINSTRUMENTS.com

MAX/MIN Function

1.  Press the 

MAX/MIN

 button to access Maximum and Minimum modes.

2.  When pressing the 

MAX/MIN

 button once, the MAX Indicator will 

appear on the display. The maximum temperature and maximum 

humidity values are displayed and will automatically update when 

new maximum data values are measured.

3.  Press the 

MAX/MIN

 button again and the MIN indicator appears on 

the display. The minimum temperature and minimum humidity values 

are displayed and will automatically update when new minimum data 

values are measured.

4.  Press the 

MAX/MIN

 button a third time and both the MAX and MIN 

indicators will blink continuously. The Maximum & Minimum values 

will automatically update in the Max/Min memory while the current 

temperature and humidity measurements are displayed. To view 

the Max/Min measurements during this mode press the 

MAX/MIN

 

button at any time to toggle through stored measurements.

5.  To exit MAX/MIN mode and resume normal operation, press and hold 

the 

MAX/MIN

 button for two seconds.

Note:

 When in MAX/MIN mode, the REL and °C/°F buttons are disabled.

Auto Power Off

To preserve battery life, the meter is programmed to turn itself off after 

30 minutes of inactivity. To disable this feature, press and hold the 

HOLD

 

button while powering on the meter. When the meter turns on, release both 

buttons simultaneously. The Auto Power Off indicator will not appear on 

the display, confirming this feature is now disabled. If the meter is turned 
off then back on, the "Auto Power Off" feature will be enabled again.

Humidity Calibration Procedure

In order to calibrate the meter humidity calibration standards are required 

(REED R9933, R9975 or R9980). If the meter has been exposed to 

extreme conditions or it has not been used in a long period of time the 
meter should be calibrated by following steps 1 through 6 below.
1.  Turn the meter off.
2.  Press the HOLD, °C/°F, and power buttons simultaneously and release.

continued...

1.800.561.8187 

[email protected]

REED-Direct.ca

Summary of Contents for R6001

Page 1: ...www REEDINSTRUMENTS com Thermo Hygrometer Instruction Manual R6001 1 800 561 8187 info REED Direct ca REED Direct ca ...

Page 2: ...umidity Calibration Procedure 6 7 Restoring Factory Settings 7 Battery Replacement 7 Accessories and Replacement Parts 7 Applications 8 Product Care 8 Product Warranty 8 Product Disposal and Recycling 9 Product Support 9 Introduction Thank you for purchasing your R6001 Thermo Hygrometer Please read the following instructions carefully before using your instrument By following the steps outlined in...

Page 3: ...or the purpose of replacing batteries may cause damage that will not be covered under the manufacturer s warranty Servicing should only be provided by an authorized service center Features Easy to use intuitive design Dual LCD readout displays temperature and humidity User selectable F C High Sampling Rate of 2 5 times sec Data hold Min Max and Relative functions Low battery indicator and auto shu...

Page 4: ...4 to 140 F 20 to 60 C Storage Temperature 14 to 140 F 10 to 60 C Operating Humidity Range 10 to 80 Dimensions 9 5 x 2 1 x 1 3 240 x 54 x 34mm Weight 6 4oz 180g Included Carrying Case Battery Instrument Description 1 Sensor Protective Cap 2 LCD Display 3 Power Button 4 MAX MIN Button 5 Data Hold Button 6 Relative Function Button 7 Unit of Measure Button 8 Battery Cover 1 2 3 4 5 6 7 8 1 800 561 818...

Page 5: ...FF Unit of Measure Selection Temperature When the meter is powered on press the C F button to toggle between Celsius and Fahrenheit Note The default unit of measure is Celsius C The meter will automati cally save the last selected unit of measure when turned off Data Hold Press the HOLD button to freeze the current reading on the display Press the button again to resume normal operation Note When ...

Page 6: ... through stored measurements 5 To exit MAX MIN mode and resume normal operation press and hold the MAX MIN button for two seconds Note When in MAX MIN mode the REL and C F buttons are disabled Auto Power Off To preserve battery life the meter is programmed to turn itself off after 30 minutes of inactivity To disable this feature press and hold the HOLD button while powering on the meter When the m...

Page 7: ...steps can be cancelled at any point by pressing the ON OFF button Restoring Factory Settings 1 Turn the meter off 2 Press the HOLD C F and power button simultaneously then release 3 When the LCD flashes press the MAX MIN button to restore factory default settings 4 A long beep will confirm factory settings have been restored The meter will then resume normal operation Battery Replacement When the ...

Page 8: ...free of defects in material or workmanship for a period of one 1 year from date of shipment During the warranty period REED Instruments will repair or replace at no charge products or parts of a product that proves to be defective because of improper material or workmanship under normal use and maintenance REED Instruments total liability is limited to repair or replacement of the product REED Ins...

Page 9: ...e contact your authorized REED distributor or REED Instruments Customer Service by phone at 1 877 849 2127 or by email at info reedinstruments com Please visit www REEDINSTRUMENTS com for the most up to date manuals datasheets product guides and software Product specifications subject to change without notice All rights reserved Any unauthorized copying or reproduction of this manual is strictly p...

Page 10: ...www REEDINSTRUMENTS com TEST MEASURE WITH CONFIDENCE CHECK OUT OUR LATEST PRODUCTS 1 800 561 8187 info REED Direct ca REED Direct ca ...

Reviews: