Measuring

Position the tip of the sensor probe within 1/4 inch (6 mm) of the  
suspected leak source. Slowly move the sensor probe past each possible 
leakage point. When the unit detects a leak source an audible tone will 
sound and visual indicators will light on the LED leak indicator display, 
appearing from left to right; green to orange to red (red being the highest 
concentration, green the lowest). Increasing levels indicates the proximity 
of the leak location is getting closer. When the unit signals a leakage, pull 
the probe away from the leak for a moment and then bring it back slowly 
to pinpoint the location. If the refrigerant leak is large setting the sensitivity 
switch to LOW (see sensitivity adjustment) will make it easier to find the 
exact site of the leak. Return the sensitivity switch to HIGH before search-
ing for any additional leaks. When you’ve finished testing for leaks turn the 
unit OFF and store it in a cool, dry, clean place to protect the sensor and 
unit from any possible damage. 

 
Ambient Concentration Reset

This combustible leak detector features an ambient concentration reset 
function that sets the unit sensors to ignore the ambient concentrations of 
refrigerant present. 

Automatic Ambient Setup

 

When turned ON the unit’s sensor will set itself to ignore the level of refriger-
ant present at the tip. Only a level or concentration great than this will cause 
the unit to alarm.

Ambient Reset Feature

 

Resetting the unit during operation performs a similar function, it programs 
the circuit to ignore the level of refrigerant present at the tip of the sensor. 
This allows the user to “home-in” on the source of the leak (higher concen-
tration). Similarly, the unit can be moved to fresh air and reset for maximum 
sensitivity. Resetting the unit with no refrigerant present (fresh air) causes 
any level above zero to be detected.

Note:

 

Be aware that this feature will cause the unit to ignore any  

refrigerant present when powered ON. If the unit sensor is pointing  
towards a leak when you turn the unit ON the alarm will not sound.

1.888.610.7664 

[email protected]

REED-Direct.com

Summary of Contents for C-380

Page 1: ...nts www com Model C 380 Refrigerant Leak Detector Instruction Manual Black Pantone 534 Blue 100 80 30 5 Pantone 485 Red 10 100 100 5 Pantone 123 Yellow 0 27 100 1 888 610 7664 info REED Direct com REE...

Page 2: ...0 5 Pantone 485 Red 10 100 100 5 Table of Contents Safety 3 Features 3 Specifications 4 Instrument Description 5 Operating Instructions 6 8 Power Up Reset 6 Testing the Unit 6 Measuring 6 Ambient Conc...

Page 3: ...nsure your safety and the proper use of this unit Features Detects and pinpoints even the smallest leaks of refrigerant gases An increasing tick rate sounds as a leak is pinpointed and the red indicat...

Page 4: ...Hi 6g year L 30g year R404A 407C 410A Hi 8g year L 40g year Alarm Settings Buzzer Tricolour LED bar indicator Sensor Advanced ionization detection Response Time Less than 1 second Warm up Time 10 seco...

Page 5: ...80 30 5 Only if you REALLY need them Pantone 534 Blue 100 80 30 5 Pantone 485 Red 10 100 100 5 Instrument Description 1 Sensor 2 Sensor Protector 3 LED Leak Indicators 4 Low Sensitivity Button 5 High...

Page 6: ...serves as an ambient concentration reset button see ambient concentration reset for more information Press ON OFF once to turn on the refrigerant leak detector The LED leak indicators will then illumi...

Page 7: ...ity switch to HIGH before search ing for any additional leaks When you ve finished testing for leaks turn the unit OFF and store it in a cool dry clean place to protect the sensor and unit from any po...

Page 8: ...cally This unit s sensor has a limited operative period and under normal operation it should stay operative for over a year To prolong the sensor s life cycle ensure that the sensor s surface is free...

Page 9: ...attery indicator will illuminate The batteries should be replaced as quickly as possible 1 Loosen the battery cover screw and remove the battery cover located on the bottom of the instrument as shown...

Page 10: ..._____________________________________ ________________________________________________ ________________________________________________ ________________________________________________ _______________...

Page 11: ..._____________________________________ ________________________________________________ ________________________________________________ ________________________________________________ _______________...

Page 12: ..._____________________________________ ________________________________________________ ________________________________________________ ________________________________________________ _______________...

Reviews: