Data Logging 

Air Quality 

Meter

Instruction 

Manual

R9910SD

 

REED Instruments

1-877-849-2127 | [email protected] | www.reedinstruments.com

Summary of Contents for R9910SD

Page 1: ...Data Logging Air Quality Meter Instruction Manual R9910SD REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...

Page 2: ...O2 Temperature Measurement Mode 10 CO2 Temperature Measurement Mode 11 Data Hold 11 Recording Maximum and Minimum Readings 11 Backlight 11 Setup Mode 12 16 Formatting the SD Card Sd F 12 13 Setting th...

Page 3: ...mm Terminal to RS 232 Serial Connection 22 System Reset 23 Battery Replacement 23 Applications 23 Accessories and Replacement Parts 24 Frequently Asked Questions FAQ s 24 25 Product Care 25 Product Wa...

Page 4: ...nder the manufacturer s warranty Servicing should only be provided by an authorized service center Features Measures oxygen O2 carbon dioxide CO2 carbon monoxide CO temperature and humidity Records up...

Page 5: ...nge 5 to 95 RH Accuracy 70 RH 3 rdg 1 RH 70 RH 3 RH Resolution 0 1 RH CO2 Measuring Range 0 to 4 000 ppm Accuracy 999 ppm 40 ppm 1 000 ppm 3000 ppm 5 rdg 3 000 ppm 250 ppm Resolution 1 ppm CO Measurin...

Page 6: ...fter 10 minutes off Kick Stand Yes Tripod Mountable Yes Low Battery Indicator Yes Power Supply 6 x AA batteries or AC Adapter optional Data Output Yes RS 232 Product Certifications CE Operating Temper...

Page 7: ...C Button 7 Rec Enter Button 8 Down Function Button 9 Log Set Sampling Check Button 10 RS 232 Output Jack 11 Reset Pin 12 Power Adapter Input Jack 13 Tripod Mounting Screw 14 Battery Cover 15 Kickstand...

Page 8: ...ta Logging Status Indicator 5 SD Card Status Indicator 6 Temperature Unit of Measure 7 CO Humidity Dew Point Units of Measure 8 Measurement Reading Indicator 9 CO2 Sensor Signal 10 Maximum and Minimum...

Page 9: ...idity CO2 and either the O2 or the CO Probe The meter will data log all three values however only one value can be displayed at a time on the LCD screen 1 Connect the Humidity CO2 and either the O2 or...

Page 10: ...rement Modes for details 2 The LCD display will indicate the wet bulb value as shown Note The wet bulb value is automatically calculated from the humidity temperature measurement CO Temperature Measur...

Page 11: ...the REC button to enter recording mode as indicated by RECORD on the LCD The meter will now begin recording maximum and minimum readings 2 While in recording mode A Press the REC button once and the...

Page 12: ...t alarm value HIgh Set CO2 height compensation value in meters CO2 mode only HIghf Set CO2 height compensation value in feet CO2 mode only ESC Exit Setup mode 3 Once the appropriate parameter has been...

Page 13: ...and skip to next parameter Note At any time you can press the ESC button to exit the Setup mode and resume normal operation The internal clock will keep accurate time when the meter is powered off Whe...

Page 14: ...ress the ENTER button to confirm selection 3 Press the SET button to save and skip to the next parameter Note At any time you can press the ESC button to exit the Setup mode and resume normal operatio...

Page 15: ...p mode and resume normal operation Advanced Setup Mode CO2 mode only Setting CO2 Height Compensation Value m Follow steps 1 through 3 when HIgh appears on the LCD 1 Press the and buttons to adjust the...

Page 16: ...mory card between 1Gb and 16Gb Insert the SD memory card in the slot at the bottom of the meter The card must be inserted with the label side facing the rear of the meter Manual Data Logging Mode In m...

Page 17: ...tton and RECORD will appear on the LCD 2 Press the LOG button to enter automatic data logging function as indicated by DATARECORD Note If a card is not inserted or the card is defective the meter will...

Page 18: ...aph in Excel See sample below 0 10 20 30 40 50 60 10 14 13 10 14 15 10 14 17 10 14 19 10 14 21 10 14 23 10 14 26 10 14 28 10 14 30 10 14 32 10 14 34 10 14 36 19 07 10 19 07 10 19 07 10 19 07 10 19 07...

Page 19: ...2 WET C 4 19 07 10 10 14 19 15 WET C 5 19 07 10 10 14 21 20 WET C 6 19 07 10 10 14 23 25 WET C 7 19 07 10 10 14 26 2 WET C 8 19 07 10 10 14 28 20 WET C 9 19 07 10 10 14 30 25 WET C 10 19 07 10 10 14...

Page 20: ...10 14 36 19 07 10 19 07 10 19 07 10 19 07 10 19 07 10 19 07 10 19 07 10 19 07 10 19 07 10 19 07 10 19 07 10 19 07 10 1 2 3 4 5 6 7 8 9 10 11 12 Ch2_CO2 CO2_unit See sample below for a CO Temperature m...

Page 21: ...first use and when not in use for extended periods Follow steps 1 through 6 if the O2 appears to be providing false readings 1 Place the probe in the air and wait a moment until the displayed reading...

Page 22: ...2 data probe 3 7 Temp O2 data probe 3 8 CO data probe 3 7 Temp CO data probe 3 8 4 47060100000209 Donn es de O2 sonde 3 20 9 O2 Lecture Position d cimale 0 3 Positif 0 Negatif 1 C 01 F 02 PPM 19 O2 06...

Page 23: ...ibutor or authorized Service Center for repair Battery Replacement When the low battery icon appears on the LCD the batteries must be replaced 1 Remove the two 2 Phillips screws on the back of the met...

Page 24: ...nts com Frequently Asked Questions FAQ s I cannot turn off my meter do you know why Often times the meter cannot be turned off because it is in recording mode REC mode Make sure to exit that function...

Page 25: ...generate a folder named AQA01 Within this folder a new file named AQA01001 XLS is created when the data logger begins recording This file is used until the Data has reached to 30 000 rows At that poi...

Page 26: ...2127 or info reedinstruments com to discuss the claim and determine the appropriate steps to process the warranty Product Disposal and Recycling Please follow local laws and regulations when disposin...

Page 27: ...Test Measure with Confidence Check out our latest Products REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...

Page 28: ...REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...