8
Operating Instructions
1. Insert the ultrasonic sensor into the probe sockets on the meter.
2. Press, hold an release the
button for 2 seconds to turn the meter
on. Press the
button to turn the meter off.
3. The LCD will briefly display information about the meter, and then
show the current set sound velocity.
Setting Ultrasonic Sensor Frequency
1. Press the
button to highlight the ultrasonic sensor frequency
setting on the display.
2. Continuously press the
button to toggle between 2MHz, 5MHz,
7MHz, 10MHz and ZW frequencies.
3. To exit, press the
button or by carrying out a measurement.
Zero Calibration
1. After setting the correct ultrasonic sensor frequency, set the sound
velocity to 5900m/s (see
Adjusting the Sound Velocity
for details).
2. Select the proper receiving gain (see
System Setup
for details).
3. Coat the 4mm calibration test block with coupling gel (R7950) and
place the sensor on the calibration test block while ensuring the
coupling indicator "
" appears on the LCD display.
4. Once the coupling indicator appears on the LCD display press the
button to initiate zero calibration.
5.
The meter will beep and then the screen will indicate that the calibration
is complete.
6. If zero calibration is not properly completed the meter will retain the
original value.
7. To delete the calibration data, see
Memory Manager
section for details.
Adjusting the Sound Velocity
1. While the meter is on, press the
button and the meter will
display the current sound velocity.
2. Press the
button to switch between the 5 pre-loaded sound
velocities, or press the
and
buttons to match the sound
velocity of the material under test as shown in the following table.
continued...