Maximum Measurement Mode

1.

Press the 

MAX

 button to enter the maximum value mode.

2.

The Max Indicator will appear on the display.

3.

Hold the durometer vertically with the point of the indenter at least
12mm from any edge.

4.

Place the presser foot onto the test piece while keeping the foot
parallel to the surface.

5.

Apply enough force to ensure proper contact between the presser
foot and the test piece.

6.

Keep in place and Hold for approx. 2 seconds.

7.

The LCD will display the measured maximum value and will hold
this value.

8.

To take the next maximum value, press the 

ZERO

 button to clear

the reading.

9.

Repeat steps 3 through 8 for each maximum value measurement.

10. Press the 

MAX

 button to exit maximum measurement mode and

resume normal operation.

Average Measurement Mode

The average value is calculated using the number of test measurements 
the user selects which can range from 1 to 9.
1.

To set the amount of test measurements required during a test
session, hold and release the 

 button to enter the average

measurement mode.

2.

The display will indicate "N" confirming it is now in average
measurement mode as well as the number of set test measurements.

3.

Press the

 button again to scroll through the required number of 

test measurements ranging from 1 to 9.

4.

Press the 

ZERO

 button to confirm the selected number of

test measurements.

5.

Press the 

MAX

 button to begin.

6.

Hold the durometer vertically with the point of the indenter at least
12mm from any edge.

continued...

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for 800837000651

Page 1: ...Scale Durometer Instruction Manual HT 6510A www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 2: ...oint Calibration 8 Battery Replacement 9 Accessories and Replacement Parts 9 Product Care 9 Product Warranty 9 Product Disposal and Recycling 10 Product Support 10 Introduction Thank you for purchasin...

Page 3: ...e of replacing batteries may cause damage that will not be covered under the manufacturer s warranty Servicing should only be provided by an authorized service center Features Shore A hardness tester...

Page 4: ...ons Yes Power Supply 2 x AAA Batteries Low Battery Indicator Yes Product Certifications CE Operating Temperature 32 to 104 F 0 to 40 C Storage Temperature 4 to 140 F 20 to 60 C Operating Humidity Rang...

Page 5: ...0A S H O RE A HARDNESS TES T E R 1 Battery Cover 2 Zero Button 3 Max Hold Button 4 LCD Display 5 Average Measaurement Button 6 POWER Button 7 Presser Foot 8 Indenter www GlobalTestSupply com Find Qual...

Page 6: ...o check accuracy simply insert the indenter into the hole of the calibrated test block 3 Apply enough force to make enough contact between the top surface of the test block and the presser foot 4 The...

Page 7: ...MAX button to exit maximum measurement mode and resume normal operation Average Measurement Mode The average value is calculated using the number of test measurements the user selects which can range...

Page 8: ...med to turn itself off after 10 minutes of inactivity Zero Calibration Press the ZERO button to perform a zero calibration when required Single Point Calibration 1 Place the indenter onto a flat test...

Page 9: ...l parts only Product Warranty REED Instruments guarantees this instrument to be free of defects in material or workmanship for a period of one 1 year from date of shipment During the warranty period R...

Page 10: ...D distributor or REED Instruments Customer Service by phone at 1 877 849 2127 or by email at info reedinstruments com Please visit www REEDINSTRUMENTS com for the most up to date manuals datasheets pr...

Page 11: ...TEST MEASURE WITH CONFIDENCE CHECK OUT OUR LATEST PRODUCTS www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 12: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Reviews: