Mounting Redsense

A flashing green indicator light indicates that the alarm unit is performing a self-test. During this test, a number of 
parameters are checked that may eventually need to be adjusted. These are displayed when the following lights 
and symbols appear:
•   The yellow indicator light, which is marked with a warning symbol, can indicate any of the following: Moisture  
  has come in contact with the sensor. The sensor has come loose, is not connected or has been broken. The
  optical extension fiber is not connected correctly.
•   If all of the indictor lights are flashing, an internal error has been detected. Contact Redsense Medical for
  additional support.
Any problems must be corrected before attempting to restart the alarm unit. If blood or moisture has come in 
contact with the sensor, it must be replaced with a new sensor. If some other technical problem occurs, do not 
use the alarm unit, rather switch to another Redsense alarm unit and contact Redsense Medical for
additional support.

During dialysis

While the unit is running, the green indicator light will light up continuously, and in the event of an error the 
following alarm signals may be heard:
•   The red indicator lights followed by a continuous alarm signal, indicate that blood has been detected by the
  sensor, either because the venous needle has been dislodged or leakage has occurred. First, deal with the leak
  and then turn off the alarm by holding in the button for at least 1 second.
•  The yellow warning indicator light followed by an intermittent alarm signal indicates one of the following:  

 

  Moisture has come in contact with the sensor. The sensor has come loose, is not connected or has been  

 

  broken. The optical extension fiber is not connected correctly. Turn off the alarm by keeping the on/off button  
  pressed, fix the cause of the alarm and then restart the alarm unit.
•   The yellow warning indicator together with the battery indicator, followed by an intermittent alarm signal,  

 

  means that the alarm unit is no longer receiving power from the AC adapter, which means that monitoring can 
  no longer be performed correctly. In this case, turn off the alarm unit by keeping the button pressed in, then    
  fix the cause of the AC power failure for the alarm unit.
•   If all the indicator lights are flashing, followed by an intermittent alarm signal, this indicates an internal error.    
  Turn off the alarm unit by holding in the button and continue running the dialysis without monitoring with    
  Redsense. Contact Redsense Medical for additional support..

After completing dialysis

1. 

 Press the on/off button for at least 1 second to turn off the alarm.

2. 

 Disconnect the sensor patch from the optical extension fiber.

3.

  Clean the optical extension fiber.

4.

  Remove the sensor patch and dispose of it according to the clinic’s procedures for handling dialysis waste.

Note: The alarm can also be turned off using the on/off button at any time during dialysis.

Maintenance

Cleaning

The actual sensor patch is made of disposable material and should not be cleaned or reused. The alarm unit and 
the optical extension fiber should be wiped off with disinfectant/alcohol and a moist cloth.

Charging the battery

The alarm unit contains a battery in order to issue an alarm in the event of a power failure from the AC adapter. 
When the alarm unit is connected to the AC adapter, the charging status for the built-in battery is displayed as:
•  A flashing battery indicator light means that the battery is being charged. The unit cannot be started in this    
  case. Wait until the charging is complete.

Summary of Contents for Alarm Unit

Page 1: ...Instructions for use Alarm unit Fiber optic cable Sensor patch ENGLISH ...

Page 2: ...edical com These instructions are released with the following date version 2017 01 18 Instructions RM 1RM240 Instructions Warnings and precautions 3 Mounting Redsense 4 During dialysis 6 After completing dialysis 6 Maintenance 6 Cleaning 6 Charging the battery 6 Replacing the battery 7 Symbols and terms 7 Technical Information 9 EMC Information 9 ...

Page 3: ...radio communication equipment may affect Redsense Redsense is not intended to be used in a highly oxygenated environment Modifying the equipment is not permitted Modification of the equipment will void all guarantees and product liability on the part of the manufacturer Servicing of the equipment may only be performed by Redsense Medical or by persons authorized by them External equipment intended...

Page 4: ... should be heard If it is the first set up or it is a new environment for the Redsense please check that the patient s supervising staff or observing person can clearly hear the alarm sound at the correct distance The alarm signal should be a continuous signal with three different levels tones and can be initiated by starting Redsense and then provoking an alarm The observing person must be sure t...

Page 5: ...5 IFU Alarm unit Fiber optic cable Sensor patch ENGLISH Warnings and precautions figure 1 figure 2 1 2 figure 3 figure 4 1 2 YES YES NO figure 5 figure 6 ...

Page 6: ...is not connected or has been broken The optical extension fiber is not connected correctly Turn off the alarm by keeping the on off button pressed fix the cause of the alarm and then restart the alarm unit The yellow warning indicator together with the battery indicator followed by an intermittent alarm signal means that the alarm unit is no longer receiving power from the AC adapter which means t...

Page 7: ...pment and changing the battery should only be done using Redsense Medical or by personnel authorized by them Symbols and terms The following symbols and terms are used in these instructions and on the product Connection for optic fiber On off button Green indicator with operating symbol Yellow indicator with battery symbol Yellow indicator with warning symbol Red indicator with alarm symbol ith Co...

Page 8: ... month YYYY MM On 0ff button Do not use if package is damaged Batchnumber includes manufacturing year Alarm indicator Item number On off indicator Serial number includes manufacturing year Warning indicator Consult the instruction for use Applied part of type BF Body Floating the surface between the equipment and the patient does not conduct electricity The alarm unit covered by the WEEE directive...

Page 9: ...recautions Alarm unit 128 x100 x 35 mm 150 g 15 to 40 C 20 to 60 C 10 to 40 C 75 75 75 5 years See EMC Information Extension fiber Ø10mm x diffent lenghts ca 10 30 g 15 to 40 C 20 to 60 C 10 to 40 C 75 75 75 5 years Sensor patch 140 x 60 x 1 mm 150 g 15 to 40 C 20 to 60 C 10 to 40 C 75 75 75 See expiry date RF emissions CISPR 11 RF emissions CISPR 11 Harmonic emission IEC 61000 3 2 Voltage fluctua...

Page 10: ...5 cycles 5 UT 95 dipinUT for 5 sec 3A m 3Vrms 3V m Electromagnetic environment Electromagnetic environment guidance Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or ho...

Page 11: ... distances between portable an mobile RF communications equipment and Redsense Redsense is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of Redsense can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and Redsense as recom m...

Page 12: ...ellow indicator with warning symbol Red indicator with alarm symbol on ator with ymbol Connector for AC adapter Service connector Redsense use only ellow indicator with attery symbol Yellow indicator with warning symbol or with ol Redsense Medical AB Gyllenhammars väg 26 302 92 HALMSTAD SWEDEN www redsensemedical com ...