Instructions for use / Alarm unit, 

Fiber optic cable & Sensor patch

ENGLISH

Summary of Contents for Alarm Unit

Page 1: ...Instructions for use Alarm unit Fiber optic cable Sensor patch ENGLISH ...

Page 2: ...edical com These instructions are released with the following date version 2017 01 18 Instructions RM 1RM240 Instructions Warnings and precautions 3 Mounting Redsense 4 During dialysis 6 After completing dialysis 6 Maintenance 6 Cleaning 6 Charging the battery 6 Replacing the battery 7 Symbols and terms 7 Technical Information 9 EMC Information 9 ...

Page 3: ...radio communication equipment may affect Redsense Redsense is not intended to be used in a highly oxygenated environment Modifying the equipment is not permitted Modification of the equipment will void all guarantees and product liability on the part of the manufacturer Servicing of the equipment may only be performed by Redsense Medical or by persons authorized by them External equipment intended...

Page 4: ... should be heard If it is the first set up or it is a new environment for the Redsense please check that the patient s supervising staff or observing person can clearly hear the alarm sound at the correct distance The alarm signal should be a continuous signal with three different levels tones and can be initiated by starting Redsense and then provoking an alarm The observing person must be sure t...

Page 5: ...5 IFU Alarm unit Fiber optic cable Sensor patch ENGLISH Warnings and precautions figure 1 figure 2 1 2 figure 3 figure 4 1 2 YES YES NO figure 5 figure 6 ...

Page 6: ...is not connected or has been broken The optical extension fiber is not connected correctly Turn off the alarm by keeping the on off button pressed fix the cause of the alarm and then restart the alarm unit The yellow warning indicator together with the battery indicator followed by an intermittent alarm signal means that the alarm unit is no longer receiving power from the AC adapter which means t...

Page 7: ...pment and changing the battery should only be done using Redsense Medical or by personnel authorized by them Symbols and terms The following symbols and terms are used in these instructions and on the product Connection for optic fiber On off button Green indicator with operating symbol Yellow indicator with battery symbol Yellow indicator with warning symbol Red indicator with alarm symbol ith Co...

Page 8: ... month YYYY MM On 0ff button Do not use if package is damaged Batchnumber includes manufacturing year Alarm indicator Item number On off indicator Serial number includes manufacturing year Warning indicator Consult the instruction for use Applied part of type BF Body Floating the surface between the equipment and the patient does not conduct electricity The alarm unit covered by the WEEE directive...

Page 9: ...recautions Alarm unit 128 x100 x 35 mm 150 g 15 to 40 C 20 to 60 C 10 to 40 C 75 75 75 5 years See EMC Information Extension fiber Ø10mm x diffent lenghts ca 10 30 g 15 to 40 C 20 to 60 C 10 to 40 C 75 75 75 5 years Sensor patch 140 x 60 x 1 mm 150 g 15 to 40 C 20 to 60 C 10 to 40 C 75 75 75 See expiry date RF emissions CISPR 11 RF emissions CISPR 11 Harmonic emission IEC 61000 3 2 Voltage fluctua...

Page 10: ...5 cycles 5 UT 95 dipinUT for 5 sec 3A m 3Vrms 3V m Electromagnetic environment Electromagnetic environment guidance Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or ho...

Page 11: ... distances between portable an mobile RF communications equipment and Redsense Redsense is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of Redsense can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and Redsense as recom m...

Page 12: ...ellow indicator with warning symbol Red indicator with alarm symbol on ator with ymbol Connector for AC adapter Service connector Redsense use only ellow indicator with attery symbol Yellow indicator with warning symbol or with ol Redsense Medical AB Gyllenhammars väg 26 302 92 HALMSTAD SWEDEN www redsensemedical com ...