10. Radio Signal Strength Input 

To enter into this mode press TEST twice and MENU once. "

rIn

" will flash on the display. This 

mode allows the Head unit and Monitor to exchange messages. The monitor knows that the Head unit is 
in this mode and resends the message. The Head unit stays awake in order to receive this resent 
message. The Head unit measures the strength of the signal and displays it (the maximum Value is 255); 
the units should work properly until the signal level drops to about 50. (

Caution

: since the radio stays 

awake for this test, the battery is being discharged much faster in this mode. Try to spend as little time 
in this test as possible.) 

 The Head unit transmits a message once per second and the Monitor responds immediately, so the test 
must be performed for at least one second. The Head unit can be moved away from obstructions, up or 
down or pointed to get the best reading. When the MENU button is pressed to leave this mode, the last 
reading taken will be stored in memory; the stored reading will be included in the messages to the 
monitor, so this reading can be examined at the Monitor. If no Monitor is in range, the readings will be 
very low or zero. "

rIn

" will flash in the display. 

11. OFF 

To put the Head unit in the

 OFF

 mode, press the TEST button for about 4 seconds. The display 

will show "

OFF

". This mode disables all transmissions by the Head unit. The unit will transmit a message 

every 5 minutes with no gas present or every 5 seconds when the gas level is equal to or greater than 
the background setting. To reduce the battery discharge, the unit can be set in 

OFF

 mode, where it will 

never transmit. To wake the Head unit, press MENU, Up or TEST, and the unit will begin the startup 
sequence. 

 

12. Startup 

When the RL101M is powered, it will display several values, like 

address

 and 

battery voltage

, then the 

unit will count up to 30 seconds and go into 

normal operation

 13. “2HI” 

If the RL 101M displays the reading “

2HI

” in normal operating mode, that means the gas level has 

earthier exceeded 5% or the sensor needs to be zeroed.  To zero, put the RL102M in “

nUL

” mode press 

the up button and it will automatically adjust the reading to zero. Then calibrate.  

For calibration see (

# 3 in basic operations

). 

Summary of Contents for RL-101M

Page 1: ...ensure it is operating properly After installation provide a copy of manual to authorized personnel IMPORTANT SAFETY WARNING Failure to follow all safety and operational instructions may result in property damage serious injury or death Power Consumption and Voltage rating RL101M instrument is powered by 6 AA batteries 9VDC internally 1mA continuous consumption when the device is operational Head ...

Page 2: ...OVERVIEW Modes 1 Normal Operation 2 NUL 3 Calibrate 4 Assign Head Address 5 Set Background Gas Level 6 Check Battery Voltage 7 Err msg 8 Revision 9 Relay Test 10 Radio Signal Strength Input 11 OFF 12 Startup 13 2HI ...

Page 3: ...engages Move the magnet repeatedly toward and away from the reed to change digits To get to the second mode this is addressing press TEST once To enter the third Mode press TEST again while in the 2nd Mode Pressing TEST while in the 3rd Mode will get back into the 2nd mode or addressing Pressing MENU multiple times will get back to Normal Operating mode If left in any mode except OFF the unit will...

Page 4: ...is in normal mode 0 00 will show on the display 2 NUL To place the RL101M into NUL Mode press Menu once nuL will flash on the display The RL101M Head unit will monitor the gas level but will not send messages when in NUL Mode NUL Mode is used for removing any errors in the amplifiers that may cause a gas reading when there is no gas Use the Up button once to automatically adjust the readings to ze...

Page 5: ...01M should automatically calibrate the sensor to 2 5 which equals the calibration gas the monitor should display 50 5 100ppm Pressing MENU will place the unit in Normal Operation and will begin sending messages to the monitor 4 Assign Radio Head to Channel To access to this feature press TEST once Adr will flash on the display when in that mode Use the UP and Down buttons to adjust the Address of ...

Page 6: ...lash on the display when in this mode 7 Error msg If the display is showing Err that typically means the sensor housing is not connected or an LEL sensor is not placed inside the sensor housing 8 Revision This mode shows the revision of the software that is being used in theRl101M Head unit To access press TEST Twice R 2 3 should show on the display 9 Relay Test To enter this mode press TEST twice...

Page 7: ... reading will be included in the messages to the monitor so this reading can be examined at the Monitor If no Monitor is in range the readings will be very low or zero rIn will flash in the display 11 OFF To put the Head unit in the OFF mode press the TEST button for about 4 seconds The display will show OFF This mode disables all transmissions by the Head unit The unit will transmit a message eve...

Page 8: ...he Rl101M Head may be mounted at a customer specified height with the sensor housing pointed down towards the ground For best signal strength position the glass window in line of sight of the Redline Monitor also make sure there are no metal nonmetal obstructions so signal will not be blocked MAINTENANCE Redline Instruments recommends you check battery voltage every 90 days and Calibration of the ...

Reviews: