background image

For Sales and Support, Contact Walker EMD • Toll-free: (800) 876-4444 • Tel: (203) 426-7700 • Fax: (203) 426-7800 • www.walkeremd.com

5

RECEIVING DATA FROM THE APLTC

Data is transmitted from the Apollo whenever a T 

(transmit command)

is

received or the hardware print request terminal is activated. The five possible
responses and a typical transmission is shown below.

The first character transmitted is the unit address number; unless it is zero, in

which case it will be left blank. Then two blank spaces. The next two characters
are the abbreviation for the values, followed by three blanks 

(two for cold

junction).

The actual data values are transmitted next. The field is right justified

with leading zeros 

(not blanks)

and negative values have a minus sign in

position 9. A plus sign for positive values is not used and note that the field for
cold junction has a decimal point. The message 

“OPEN”

will appear in the data

field of the thermocouple temperature transmission whenever the indicator
senses a thermocouple break. After the data field, an F or C character is
transmitted depending on the indicator’s current units, followed by carriage
return and line feed.

SERIAL TERMINAL DESCRIPTION

Note: The numbers in parentheses refer to the position numbers labeled on the

terminal block.

SI+

(1) - Serial In (+):

Current flows into the Apollo’s receive channel.

SI-

(2) - Serial In (-):

Current flows out of the Apollo’s receive channel.

SO+

(3) - Serial Out (+):

Current flows into the Apollo’s transmit channel.

SO-

(4) - Serial Out (-):

Current flows out of the Apollo’s transmit channel.

20 mA- (5) - 20 mA current source for either receive or transmit loop current. 

Current flows into this pin.

20 mA+(6) - 20 mA current source for either receive or transmit loop current. 

Current flows out of this pin.

When wiring the 20 mA Serial Loop, remove the 6-position terminal block

on the right side of the bottom board. Strip 1/4" off each wire and connect each
wire into the proper location of the terminal block. Re-install the connector and
be sure that the wires are strain-relieved.

ANALOG OUTPUT OPTION

The Apollo Thermocouple Indicator Analog Output Option provides an

isolated 1mV/degree output to drive chart recorders, controllers, data loggers
and slave displays. The output corresponds to the displayed temperature and
increments in discrete steps. Although the option is calibrated at the factory,
user accessible zero and span adjustments are provided for minor adjustments.
These are located adjacent to the output connections. Recalibration is
recommended after 1 year of use.

The analog output option terminal block is located at the rear of the unit in

the upper right-hand corner. Terminal 4 is the VOLT+, and terminal 5 is the
VOLT- 

(COMM.) (Note:The alarm COMM. terminal and the analog output

VOLT- terminal are internally connected.)

To check for accuracy, connect a voltmeter to the output and the reading should

be +1mV per degree of displayed reading 

(+/- 6 degrees)

. If calibration should be

necessary, connect the indicator as shown in the set-up and calibration section. 

1. Apply an arbitrary voltage so that the indicator displays a negative

temperature 

(-100,-200, etc.)

. Adjust the zero potentiometer

(right side)

so

that the readings from the voltmeter and the indicator correspond.

2. Apply an arbitrary voltage so that the indicator displays a large positive

temperature 

(+1500, +2000, etc.)

. Adjust the span potentiometer 

(left side)

so that the reading from the voltmeter and the indicator correspond.

3. Repeat steps 1 and 2 until both points are correct. Calibration is complete. 

TROUBLE SHOOTING

A functional test can be performed by connecting a wire jumper between the

thermocouple inputs. When power is applied, the display will briefly show
“888.8

8

” to test the LEDs and then indicate the cold junction temperature at the

rear of the unit. Verify that all display segments illuminate and the temperature
indicated is near ambient temperature. The indicator has several special display
modes which indicate faults within the meter or external connections. The table
illustrates these and the most common causes.

CALIBRATION

The Apollo thermocouple indicator has been calibrated at the factory using

precision instruments and was designed not to require frequent recalibration.

However, as part of a regular maintenance program or after 1 year of use,

recalibration by authorized personnel is recommended.

Recalibration consists of applying accurate voltages to the indicator and

setting the cold junction temperature with an accurate thermometer. A precision
thermometer (RTD, thermistor or similar type) (accuracy of ±0.3°C) and an
accurate voltage source (0.01%) are required. 

Do not use thermocouple wire

when making connections.

Allow 30 minutes warm-up time before calibration is attempted.
If long thermocouple runs are present in the application 

(>300 ft.),

significant

offset errors may exist due to the input bias current of the indicator. To correct
for this effect, insert a fixed resistance in series with the positive leg of the
millivolt source which is equal to the thermocouple wire resistance and proceed
with calibration as normal. In most thermocouple applications this error is
insignificant and may be ignored.

Following the procedure outlined in SET-UP and CALIBRATION, press the

buttons indicated in the following charts and observe the indicated display.

In steps D and E of CAL 

4

, a voltage applied that is not within the calibration

range of the instrument will cause the calibration process to be aborted after 20
seconds and return the display to CAL 

0

. Check the calibration set-up before

attempting to recalibrate.

In step C, of CAL 

5

, the display will first indicate what the display is and after

2 seconds indicate the current value.

In step D, of CAL 

5

, when the Select button is pressed, the display will

increment first slowly and then more rapidly. The button can be released and
pressed again to increment slowly when the desired value is approached.

NOTE: 

*

A decimal point will be 

placed in this position 
for cold junction.

1

TC

0075F

1

CJ

023.9F

Alarm high value

1

SH

0800F

Alarm low value

1

SL

-045F

Hysteresis

1

Hy

0005F

PARAMETER

MNEMONICS

DATA

Thermocouple temperature

UNIT ADDRESS

Reference junction temp.

DISPLAY

“OPEN”

“EEEE”

PROBABLE CAUSES

-   incorrect power up/down due to voltage spike

-   broken or burned-out thermocouple
-   excessive thermocouple voltage

-   excessive negative thermocouple voltage (usually due to

reversed T/C connections) defect in analog circuit

“888.88”

Or

“- - - -”

STEP

CAL. BUTTON

SEL. BUTTON

DISPLAY

A

X 10 secs.

CAL 

0

B

X

CAL 

4

C

X

In-0  Apply 0 mill/-3 µV

D

X

X

In-5  Apply 50 mill/-5 µV

E

X

X

CAL 

0

It is not necessary to exit to the thermocouple display if further set-up or calibration are

required, just press the “SEL” button. While in CAL 

0

, select the desired CAL #.

F

X

888.8

8

Then TC Disp.

*

CAL 4    ZERO AND SPAN CALIBRATION

Summary of Contents for APLTC

Page 1: ...FECT 20 V 350 7 OPEN THERMOCOUPLE INDICATION Display OPEN Analog Output 500 mV output Serial Output OPEN in data field 8 COLD JUNCTION COMPENSATION ERROR 0 02 degree degree 9 READING RATE 2 5 readings...

Page 2: ...rejection A non volatile E2PROM memory device provides permanent data retention for microprocessor variables The display consists of 5 solid state LEDs for temperature and units readout The serial com...

Page 3: ...pidly The button can be released and pressed again to increment slowly when the desired value is approached SERIAL COMMUNICATIONS OPTION DESCRIPTION The Apollo Thermocouple Indicator Serial Communicat...

Page 4: ...Apollo must perform and the identifier on which it operates 3 If the change command V is being used the next characters in the string after the value identifier are the numerical data 4 All commands...

Page 5: ...temperature 1500 2000 etc Adjust the span potentiometer left side so that the reading from the voltmeter and the indicator correspond 3 Repeat steps 1 and 2 until both points are correct Calibration i...

Page 6: ...anneal one type of alloy it must be brought to 700 F and left to soak for 2 hours after which it must cool at the rate of 100 hour until it reaches 500 where it must stay for 1 hour and then allowed t...

Page 7: ...trudes through the other side Install each of the mounting clips by inserting the wide lip of the clips into the wide end of the hole located on either side of the case Then snap the clip onto the cas...

Page 8: ...or products containing components manufactured by RLC and based upon personal injuries deaths property damage lost profits and other matters which Buyer its employees or sub contractors are or may be...

Reviews: