CMM 100 :
INSTALLATION, CALIBRATION AND OPERATION
9
2.6.2 Set High/ Low Moisture Content Alarm Threshold
Having first set the
"HIGH /
"
LO
" alarm option (section 2.6.1), you can then
individually set the alarm threshold (% M.C.) for each channel as desired.
1.
In CAL mode
press the button to cycle through the factors until "
ALAr
"
is displayed. The default threshold is displayed for the selected channel.
2.
Press the button to select the desired channel.
3.
Edit the setting as described in section 3.2.
NOTE: Changing the "
HIGH
" / "
LO
" alarm option resets any pre-programmed
threshold values to the factory default values (25% / 5%).