CALIBRATION
26
“
GrAd
”
xxxx
(no units).
Compensation Curve Factor
: This gives the correction for the change in lift
speed. If a faster lift gives a lower weight reading, then decrease the “
GrAd
”
figure.
“
ProP
”
xxxx
(%)
Proportional Correction Factor:
This gives the reduction in speed
compensation for an empty bucket, compared to a full bucket. Sometimes
after an Autocal, it is found that the speed correction with a full bucket is OK,
but with an empty bucket there is a larger error (e.g. 50 –60kg).
If with an empty bucket, the weight reading is lower at a slower lift speed than
at a fast speed, then reduce the factor by 2 – 3% until the error is removed.
“
FrEq
”
xxxx
(Hertz)
Frequency at full load (determined from Autocal):
6.3.3
Live Static (Channel 2)
This sets how the weight reading is displayed in Static Weighing mode.
“
OFF
” The weight reading is “frozen” on the display after the weight has
been sampled.
“
ON
”
You get a “live” weight reading while the bucket is at the weighing
position.
“
AUTO
” You get a live reading, but it will only change if there is a change of
5% or more in the calculated weight (default setting).
Press to toggle.