USER GUIDE
V1.0
REMOTE CARE PARTNERS
PAGE 39
Intermediate measurement precision (which
includes variability across multiple days) was
evaluated using control solutions at 3 glucose
levels. With each control solution, each of 3 lots
of Contour Next test strips was tested once on
each of 10 instruments on 10 separate days for a
total of 300 readings. The following precision
results were obtained.
Table 2: System intermediate precision results for Contour
Next ONE meter using Contour Next test strips